Open Access 20.03.2023 | Knee Arthroplasty

Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation

verfasst von: Maria Anna Smolle, Amir Koutp, Clemens Clar, Lukas Leitner, Andreas Leithner, Patrick Sadoghi

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 9/2023

Abstract

Introduction

In total knee arthroplasty (TKA), tibial obliquity-restoration using kinematic alignment (KA) poses a major difference to conventional mechanical alignment. This study aimed at analysing the accuracy of conventional instrumentation (CI) versus patient-specific instrumentation (PSI) to restore anatomic tibial obliquity measured by the medial proximal tibial angle (MPTA) on conventional X-rays.

Materials and Methods

One-hundred patients were randomized to receive CI (n = 50) or PSI (n = 50) for TKA. Further 100 patients received CI without randomisation, resulting in 200 patients in total (127 women, mean age: 70.7 (range: 48–90 years). Pre- and postoperative X-rays were measured twice by two observers with a 2-week break in-between. Inter- and intraclass correlations were calculated and postoperative tibial obliquity compared to preoperative anatomy.

Results

In 150 patients with CI, no case with tibial obliquity-deviation greater than 2° was found, whilst 21.3% (n = 32) and 0.7% (n = 1) of cases and had a deviation of 0°–1°, and 1°–2°, respectively. In the remaining 78.0% (n = 117), tibial obliquity was restored. In 50 patients with PSI, no single case with a deviation greater than 1° was found. Sixty percent (n = 30) had a deviation of 0°–1°. In the remaining 40.0% (n = 20), no deviation from preoperative measurements was found. Consequently, CI resulted in a significantly smaller change in tibial obliquity from preoperative to postoperative than PSI (p < 0.001). Inter- and intra-class correlations showed a substantial agreement (any ICC > 0.90).

Conclusion

Both conventional and patient-specific instrumentation revealed adequate results with respect to restoring tibial obliquity in kinematically aligned TKA, with conventional instrumentation achieving superior results.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

For decades, surgeons tried to achieve neutral tibial alignment with a medial proximal tibial angle (MPTA) of 90° and, therefore, were not trained to align the tibia according to this physiological MTPA equivalent to a mean of 2.9° with a range from 20.5° of valgus to 20.5° of varus [17]. The theoretical advantage of neutral mechanical alignment—with the mechanical axis passing mid-line—is the even load distribution in the medial and lateral femorotibial compartment with consecutively reduced risk for implant loosening and wear [35, 13, 16]. In literature, however, this is discussed controversially [1]. In addition, mechanical alignment does not necessarily depict patients’ individual knee anatomy. In fact, only 0.1% of TKA patients actually have neutral femoral and tibial mechanical axes, whilst the vast majority presents with some degree of variation [2, 8]. Even more, according to a study by Bellemans et al. a constitutional varus knee, defined as natural mechanical alignment of at least 3° varus, can be found in 17% and 32% of asymptomatic adult women and men, respectively [3]. Upon TKA, neutral mechanical alignment in these patients would overcorrect the pre-existing varus, and eventually require some sort of medial soft tissue release [3]. Based on these observations, the philosophy of kinematic alignment (KA) was first proposed by Howell et al. aiming at a more patient-tailored reconstruction of the mechanical axis [8, 10]. First long-term results on KA are encouraging, with implant survivorship of 97.4% at 10 years [11].
Methods to achieve KA in TKA include caliper verification (as proposed by Howell et al.), robotics, computer-aided surgery, or patient-specific instrumentation (PSI) [8, 15]. In comparison to conventional instrumentation with standard blocks, PSI uses customised cutting blocks based on 3D models deriving from preoperative magnetic resonance imaging (MRI) or computed tomography (CT) scans [14], to achieve a more individualized reconstruction of preoperative knee anatomy [17]. However, the definite decision which instrumentation should be used to achieve KA still has to be made.
The aim of the current study was, therefore, to evaluate the accuracy of conventional instrumentation (CI) versus patient-specific instrumentation (PSI) with respect to restoration of tibial obliquity (based on mechanical medial proximal tibial angle [MPTA]) measured on conventional X-rays. It was hypothesized that PSI would be superior to conventional instrumentation with respect to accuracy of restoring tibial obliquity.

Material and methods

This study was approved by the institutional review board (blinded for review). Of 386 patients undergoing TKA at a single high-volume orthopaedic surgery centre due to end-stage osteoarthritis (grade IV according to Kellgren and Lawrence) of 2 out of 3 compartments of their joint, one-hundred patients were prospectively included between May 2020 and May 2022 and randomized into CI (n = 50) or PSI (n = 50) groups. Further 100 patients from the initial cohort directly underwent TKA with CI (Fig. 1). Mean age at surgery of the entire cohort (n = 200) was 70.7 (range: 48–90) years, and 127 were women (63.5%). All procedures were performed by one experienced senior knee surgeon.

Inclusion and exclusion criteria

All patients with primary knee osteoarthritis above the age of 50 years requiring TKA were potentially eligible. Patients younger than 50 years of age, those with a history of tibial or femoral fracture, osteotomy, septic arthritis, rheumatoid arthritis, or previous (partial) knee arthroplasty at the side of osteoarthritis, were excluded (Fig. 1). Prior to study participation, all patients gave their written informed consent.

Surgical technique

All patients were operated using the medial pivot cemented GMK Sphere system (Medacta, Castel San Pietro, Switzerland). The group with CI was operated on according to the calipered technique by Howell [9]. Tibial obliquity was measured preoperatively with respect to the MPTA on X-ray. After finishing the femur with the calipered technique [9], the tibial obliquity was established using two tibial stylus with 8 mm to reference medial and lateral to the tibial spine at the border where no cartilage wear was evident to restore pre-arthritic anatomy (Fig. 2 A–E).
Intraoperatively, tibial obliquity was checked for plausibility using a tibial alignment rod in line with preoperative X-ray of the tibia (Fig. 3). Tibial recuts were performed in case of inadequate extension or flexion gaps [9].
The group with PSI was operated according to the surgical technique provided by the company (Medacta, Castel San Pietro, Switzerland) using CT-based individual cutting jigs as presented by previous study groups [12]. Preoperative planning for the PSI group was performed by the company and double checked for verification by the senior surgeon in every single case. After finishing the distal femoral cut, the tibial cut was performed to evaluate the extension gap. Next, the femur was finished using the anatomic rotation as illustrated by the PSI guide.
For all patients in both techniques, the desired balance was equally balanced in full extension and laterally lax in 90 degrees of flexion according to the rationale of kinematic alignment [911].

Radiological measurements

Tibial obliquity, defined as the MPTA, was measured by two independent investigators on pre- and postoperative x-rays of the entire limb using mediCAD 2D® (Hectec GmbH, Germany). MPTA was defined as the medial angle formed between the tangent to the tibial plateau line and the tibial mechanical axis. Every investigator measured images twice in a random fashion. The second round of measurements took place at least 2 weeks after the first one. Drop-out rate of this study was 0%.

Statistical analysis

For continuous variables with parametric and non-parametric distribution, means (with standard deviations [SDs]) and medians (with interquartile ranges [IQR]) were calculated. Definite values of pre- and postoperative tibial obliquity were calculated as means of the two mean measurements obtained by the two investigators. T-tests were used to assess differences in continuous variables between groups. To assess changes in continuous variables for more than two groups, Kruskal–Wallis tests were used. Changes in tibial obliquity (based on the MPTA) from pre- to postoperative were grouped into I. restoration of tibial obliquity, II. change 0°–1°, III. change 1°–2°, and IV. change > 2°.
Differences in tibial obliquity depending on treatment group were compared with chi-squared tests. Intraclass correlation coefficients (ICCs) between the reviewers were calculated for the two measurements. A-priori sample size calculation with a p-value of < 0.05 and a power greater than 80% revealed n = 50 cases per group as sufficient to detect a clinically relevant difference of 10% of deviation in the range from 0 to 1°, or above. A p-value of < 0.05 was considered statistically significant.

Results

There were no significant differences in terms of gender distribution (p = 0.932), age at surgery (p = 0.290) or BMI (p = 0.141) between the two treatment groups (Table 1).
Table 1
Characteristics of the study population split by type of surgery
  
Entire cohort (n = 200)
Conventional instrumentation (n = 150)
PSI (n = 50)
P value
Gender
Male
73 (36.5)
55 (36.7)
18 (36.0)
0.932
Female
127 (63.5)
95 (63.3)
32 (64.0)
Age at surgery (in years, mean ± SD)
70.7 ± 9.6
70.3 ± 9.5
71.9 ± 9.8
0.290
BMI (mean ± SD)
30.0 ± 5.3
30.3 ± 5.5
29.1 ± 4.6
0.141
Mean preoperative tibial obliquity amounted to 89.9° (range: 86°–93°). There was a significant difference in preoperative tibial obliquity depending on treatment group, with greater varus tibial obliquity in the CI (mean: 89.6° [range. 86°–93°]) than PSI group (mean: 90.6° [range: 88°–92°]; p < 0.001).
For the entire cohort, the postoperative tibial obliquity amounted to 89.9° (range: 85°–93°). There was a significant difference in postoperative tibial obliquity between treatment groups, again with greater varus obliquity in the CI (mean: 89.7° [range: 85°–93°]) than PSI group (mean: 90.5° [range: 88°–92°]; p < 0.001).
Overall, no change in tibial obliquity from pre- to postoperative larger than 2° was found. One patient presented with change in tibial obliquity between 1° and 2° (0.5%), and 62 with a change between 0° and 1° (31.0%). In the remaining 137 patients, preoperative tibial obliquity had been restored, with no deviation from preoperative measurement (68.5%).
Notably, in a significantly higher proportion of patients with CI, preoperative tibial obliquity had been restored (78.0%; n = 117) as compared to PSI (40.0%; n = 20; p < 0.001). Furthermore, 21.3% (n = 32) and 60.0% (n = 30) of patients with CI and PSI, respectively, presented with a tibial obliquity change between 0° and 1°. One change in tibial obliquity between 1° and 2° was observed with the conventional technique (0.7%). In detail, PSI rather resulted in valgus deviation from preoperative tibial obliquity than CI, i.e., a mean increase in tibial obliquity of 0.1° (range: − 1°–0.5°) was observed for PSI, compared to a mean decrease of − 0.1° (range: − 1.5°–1°) for CI. Moreover, change in tibial obliquity was comparable between men and women (p = 0.642), and did not alter significantly with increasing age (p = 0.169), or BMI (p = 0.320).
ICCs between reviewers for pre- (ICC = 0.988) and postoperative measurements (ICC = 0.999) showed excellent agreement.

Discussion

According to the present study, both PSI and conventional instrumentation achieved adequate results with respect to restoring tibial obliquity in kinematically aligned TKA, and that CI was superior to PSI.
Till now, no study comparing KA with PSI and CI regarding change in tibial obliquity has been published in literature. However, some studies have focused on change in tibial obliquity comparing PSI-based kinematic with conventional mechanical alignment [6, 7, 18]. Radiological results of the present study (no change in tibial obliquity > 2°) are, for example, better than the ones observed by Waterson et al. in a in a prospective randomized controlled trial comparing KA with PSI (n = 71) vs. conventional mechanical alignment (n = 71) using the Stryker Triathlon (Stryker Navigation, Kalamazoo, Michigan, US) system [18]. They reported on 78% and 77% of kinematically and mechanically aligned TKAs to be within 3° of preoperative alignment, but did not provide statistical tests [18]. Other than in the present study, Waterson et al. referred to alignment measurements based on preoperative MRI scans that had been performed in the PSI group [18], whereas radiological alignment measurements of the present study were based on x-rays.
Calliess et al. performed a similar prospective randomized controlled trial in 100 patients with PSI and conventional mechanical instrumentation [6]. Comparable to our observations, they reported on larger deviations in postoperative limb alignment following PSI than conventional mechanical instrumentation, although not providing statistical tests [6].
Dossett et al. likewise carried out a prospective randomized controlled trial comparing KA with PSI (n = 44) and conventional mechanical alignment (n = 44) using the Vanguard (Biomet Inc., Warsaw, Indiana, US) system [7]. In their study, knee and limb alignment was comparable between groups, whilst knee joint alignment of patients in the kinematic group was significantly more valgus (by 1.9°), the tibial component significantly more varus (by 2.1°), and the femoral component significantly more valgus (by 2.2°) compared to the mechanically aligned group [7]. This corroborates our findings, with PSI leading to tibial obliquity in more valgus than CI, although we used CI with kinematic rather than mechanical alignment.
A limitation of the current study is the lack of clinical outcome data (e.g., WOMAC, KSS) to correlate with radiographic findings made. Furthermore, 3 patients had to be excluded secondarily due to missing postoperative knee images. However, the resulting drop-out rate of 0.5% may be considered negligible. Also, tibial obliquity was measured on a.p. x-rays only, whereas the sagittal alignment (i.e., tibial slope) had not been considered in the current study.
Furthermore, 50 patients were randomized into both groups. This process was further randomized due to visiting surgeons requesting a specific technique. In addition, 100 patients were operated without randomization on using the CI instrumentation due to lack of available CT scans for PSI.

Conclusions

Both conventional and patient-specific instrumentation revealed adequate results with respect to restoring tibial obliquity in kinematically aligned TKA, with conventional instrumentation presenting superior results. In the future, the conventional technique may be preferred to PSI due to fewer radiation exposure and reduced costs. Nevertheless, further studies on larger cohorts might reveal the effect on potential outlier limitation using PSI.

Declarations

Conflict of interest

A.L. has received institutional educational grants by Johnson & Johnson, Alphamed, Implantec, Medacta. P.S. is an editorial board member of The Journal of Arthroplasty, as well as KSSTA and Arthroscopy. The remaining co-authors have no conflicts of interest to declare.

Ethical approval

The study has been approved by the local institutional review board (IRB-number: 31–176 ex 18/19).
All patients gave their written informed consent prior to study participation.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Abdel MP, Ollivier M, Parratte S, Trousdale RT, Berry DJ, Pagnano MW (2018) Effect of postoperative mechanical axis alignment on survival and functional outcomes of modern total knee arthroplasties with cement: a concise follow-up at 20 years. J Bone Joint Surg Am 100(6):472–478CrossRefPubMed Abdel MP, Ollivier M, Parratte S, Trousdale RT, Berry DJ, Pagnano MW (2018) Effect of postoperative mechanical axis alignment on survival and functional outcomes of modern total knee arthroplasties with cement: a concise follow-up at 20 years. J Bone Joint Surg Am 100(6):472–478CrossRefPubMed
2.
Zurück zum Zitat Almaawi AM, Hutt JRB, Masse V, Lavigne M, Vendittoli PA (2017) The impact of mechanical and restricted kinematic alignment on knee anatomy in total knee arthroplasty. J Arthroplasty 32(7):2133–2140CrossRefPubMed Almaawi AM, Hutt JRB, Masse V, Lavigne M, Vendittoli PA (2017) The impact of mechanical and restricted kinematic alignment on knee anatomy in total knee arthroplasty. J Arthroplasty 32(7):2133–2140CrossRefPubMed
3.
Zurück zum Zitat Bellemans J, Colyn W, Vandenneucker H, Victor J (2012) The Chitranjan ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res 470(1):45–53CrossRefPubMed Bellemans J, Colyn W, Vandenneucker H, Victor J (2012) The Chitranjan ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res 470(1):45–53CrossRefPubMed
4.
Zurück zum Zitat Benjamin J (2006) Component alignment in total knee arthroplasty. Instr Course Lect 55:405–412PubMed Benjamin J (2006) Component alignment in total knee arthroplasty. Instr Course Lect 55:405–412PubMed
5.
Zurück zum Zitat Berend ME, Ritter MA, Meding JB, et al. (2004) Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res(428):26–34 Berend ME, Ritter MA, Meding JB, et al. (2004) Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res(428):26–34
6.
Zurück zum Zitat Calliess T, Bauer K, Stukenborg-Colsman C, Windhagen H, Budde S, Ettinger M (2017) PSI kinematic versus non-PSI mechanical alignment in total knee arthroplasty: a prospective, randomized study. Knee Surg Sports Traumatol Arthrosc 25(6):1743–1748CrossRefPubMed Calliess T, Bauer K, Stukenborg-Colsman C, Windhagen H, Budde S, Ettinger M (2017) PSI kinematic versus non-PSI mechanical alignment in total knee arthroplasty: a prospective, randomized study. Knee Surg Sports Traumatol Arthrosc 25(6):1743–1748CrossRefPubMed
7.
Zurück zum Zitat Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG (2014) A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Joint J 96(7):907–913CrossRefPubMed Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG (2014) A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Joint J 96(7):907–913CrossRefPubMed
8.
Zurück zum Zitat Hirschmann MT, Becker R, Tandogan R, Vendittoli PA, Howell S (2019) Alignment in TKA: what has been clear is not anymore! Knee Surg Sports Traumatol Arthrosc 27(7):2037–2039CrossRefPubMed Hirschmann MT, Becker R, Tandogan R, Vendittoli PA, Howell S (2019) Alignment in TKA: what has been clear is not anymore! Knee Surg Sports Traumatol Arthrosc 27(7):2037–2039CrossRefPubMed
9.
Zurück zum Zitat Howell SM (2019) Calipered kinematically aligned total knee arthroplasty: an accurate technique that improves patient outcomes and implant survival. Orthopedics 42(3):126–135CrossRefPubMed Howell SM (2019) Calipered kinematically aligned total knee arthroplasty: an accurate technique that improves patient outcomes and implant survival. Orthopedics 42(3):126–135CrossRefPubMed
10.
Zurück zum Zitat Howell SM, Howell SJ, Kuznik KT, Cohen J, Hull ML (2013) Does a kinematically aligned total knee arthroplasty restore function without failure regardless of alignment category? Clin Orthop Relat Res 471(3):1000–1007CrossRefPubMed Howell SM, Howell SJ, Kuznik KT, Cohen J, Hull ML (2013) Does a kinematically aligned total knee arthroplasty restore function without failure regardless of alignment category? Clin Orthop Relat Res 471(3):1000–1007CrossRefPubMed
11.
Zurück zum Zitat Howell SM, Shelton TJ, Hull ML (2018) Implant survival and function ten years after kinematically aligned total knee arthroplasty. J Arthroplasty 33(12):3678–3684CrossRefPubMed Howell SM, Shelton TJ, Hull ML (2018) Implant survival and function ten years after kinematically aligned total knee arthroplasty. J Arthroplasty 33(12):3678–3684CrossRefPubMed
12.
Zurück zum Zitat Koch PP, Müller D, Pisan M, Fucentese SF (2013) Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique. Knee Surg Sports Traumatol Arthrosc 21(10):2200–2205CrossRefPubMed Koch PP, Müller D, Pisan M, Fucentese SF (2013) Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique. Knee Surg Sports Traumatol Arthrosc 21(10):2200–2205CrossRefPubMed
13.
Zurück zum Zitat Lotke PA, Ecker ML (1977) Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRefPubMed Lotke PA, Ecker ML (1977) Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRefPubMed
14.
Zurück zum Zitat Mattei L, Pellegrino P, Calò M, Bistolfi A, Castoldi F (2016) Patient specific instrumentation in total knee arthroplasty: a state of the art. Ann Transl Med 4(7):126CrossRefPubMedPubMedCentral Mattei L, Pellegrino P, Calò M, Bistolfi A, Castoldi F (2016) Patient specific instrumentation in total knee arthroplasty: a state of the art. Ann Transl Med 4(7):126CrossRefPubMedPubMedCentral
15.
16.
Zurück zum Zitat Park JK, Seon JK, Cho KJ, Lee NH, Song EK (2018) Is Immediate postoperative mechanical axis associated with the revision rate of primary total knee arthroplasty? a 10-year follow-up study. Clin Orthop Surg 10(2):167–173CrossRefPubMedPubMedCentral Park JK, Seon JK, Cho KJ, Lee NH, Song EK (2018) Is Immediate postoperative mechanical axis associated with the revision rate of primary total knee arthroplasty? a 10-year follow-up study. Clin Orthop Surg 10(2):167–173CrossRefPubMedPubMedCentral
17.
18.
Zurück zum Zitat Waterson HB, Clement ND, Eyres KS, Mandalia VI, Toms AD (2016) The early outcome of kinematic versus mechanical alignment in total knee arthroplasty: a prospective randomised control trial. Bone Joint J 98(10):1360–1368CrossRefPubMed Waterson HB, Clement ND, Eyres KS, Mandalia VI, Toms AD (2016) The early outcome of kinematic versus mechanical alignment in total knee arthroplasty: a prospective randomised control trial. Bone Joint J 98(10):1360–1368CrossRefPubMed
Metadaten
Titel
Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation
verfasst von
Maria Anna Smolle
Amir Koutp
Clemens Clar
Lukas Leitner
Andreas Leithner
Patrick Sadoghi
Publikationsdatum
20.03.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 9/2023
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-023-04845-7

Weitere Artikel der Ausgabe 9/2023

Archives of Orthopaedic and Trauma Surgery 9/2023 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.