12.07.2023 | Hip Arthroplasty

In vivo creep and wear performance of vitamin-E-diffused highly crosslinked polyethylene in total hip arthroplasty

verfasst von: Tsunehiro Matsuoka, Yasuhito Takahashi, Tsunehito Ishida, Toshiyuki Tateiwa, Takaaki Shishido, Kengo Yamamoto

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 12/2023

Einloggen, um Zugang zu erhalten

Abstract

Introduction

An acetabular liner thickness of around 6 mm remains the “gold standard” in total hip arthroplasty. Some surgeons have been recommending the use of the thickest possible liner because contact stress and strain in articulating surfaces decrease with increasing the wall thickness. The purpose of this study was to determine whether in vivo creep and wear performance could be enhanced using a thicker liner over the standard thickness in vitamin-E-diffused highly crosslinked polyethylene (HXLPE).

Materials and methods

One hundred and twenty-two hips were allocated to age-matched, sex-matched, and body mass index-matched two subgroups implanted either with a 6.8- or 8.9-mm-thick vitamin-E-diffused HXLPE liner against 28-mm cobalt-chrome femoral head, and followed-up for 7 years. Linear and volumetric penetration of femoral head into the liners attributed to creep and wear were analyzed for each group.

Results

Compressive creep strain generated at the initial 6 months was significantly larger in the 6.8-mm group (2.6%) than in the 8.9-mm group (2.2%). The linear steady-state wear observed after 2 years was 0.0019 and 0.0015 mm/year, whereas the volumetric steady-state wear was 0.54 and 0.45 mm3/years in the 6.8- and 8.9-mm-thick groups, respectively. Although less strain in the thicker group resulted in a slightly less wear, it did not reach significant differences in the steady-state wear rates between the groups.

Conclusion

No clinical significance for using a thicker liner over the standard thickness (6.8 mm → 8.9 mm) was confirmed in the vitamin-E-diffused HXLPE according to the 7-year follow-up. The wear rates for both thicknesses were very low enough to prevent osteolysis, and no mechanical failure was observed at any follow-up interval. Nevertheless, since the significantly higher strain was seen in the thinner liner, further follow-up is needed to compare the longer term wear and the incidence of osteolysis and component fracture.
Literatur
1.
Zurück zum Zitat Khoshbin A, Wu J, Ward S, Melo LT, Schemitsch EH, Waddell JP, Atrey A (2020) Wear rates of XLPE nearly 50% lower than previously thought after adjusting for initial creep: an RCT comparing 4 bearing combinations. JB JS Open Access. 5(2):e0066CrossRefPubMedPubMedCentral Khoshbin A, Wu J, Ward S, Melo LT, Schemitsch EH, Waddell JP, Atrey A (2020) Wear rates of XLPE nearly 50% lower than previously thought after adjusting for initial creep: an RCT comparing 4 bearing combinations. JB JS Open Access. 5(2):e0066CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Nielson T, Owens G, Miller B, Meneghini E, Deckard ER, Meneghini RM (2022) Large femoral heads in total hip arthroplasty with vitamin E highly cross-linked polyethylene: head penetration rates compared to highly cross-linked polyethylene. J Arthroplast 37:S685–S691CrossRef Nielson T, Owens G, Miller B, Meneghini E, Deckard ER, Meneghini RM (2022) Large femoral heads in total hip arthroplasty with vitamin E highly cross-linked polyethylene: head penetration rates compared to highly cross-linked polyethylene. J Arthroplast 37:S685–S691CrossRef
3.
Zurück zum Zitat Rochcongar G, Buia G, Bourroux E, Dunet J, Chapus V, Hulet C (2018) Creep and wear in vitamin E-infused highly cross-linked polyethylene cups for total hip arthroplasty: a prospective randomized controlled trial. J Bone Jt Surg Am 100:107–114CrossRef Rochcongar G, Buia G, Bourroux E, Dunet J, Chapus V, Hulet C (2018) Creep and wear in vitamin E-infused highly cross-linked polyethylene cups for total hip arthroplasty: a prospective randomized controlled trial. J Bone Jt Surg Am 100:107–114CrossRef
4.
Zurück zum Zitat Matar HE, Platt SR, Board TN, Porter ML (2020) Overview of randomized controlled trials in primary total hip arthroplasty (34,020 Patients): what have we learnt? J Am Acad Orthop Surg Glob Res Rev 4(8):e20.00120PubMedPubMedCentral Matar HE, Platt SR, Board TN, Porter ML (2020) Overview of randomized controlled trials in primary total hip arthroplasty (34,020 Patients): what have we learnt? J Am Acad Orthop Surg Glob Res Rev 4(8):e20.00120PubMedPubMedCentral
5.
Zurück zum Zitat Shi J, Zhu W, Liang S, Li H, Li S (2021) Cross-linked versus conventional polyethylene for long-term clinical outcomes after total hip arthroplasty: a systematic review and meta-analysis. J Investig Surg 34(3):307–317CrossRef Shi J, Zhu W, Liang S, Li H, Li S (2021) Cross-linked versus conventional polyethylene for long-term clinical outcomes after total hip arthroplasty: a systematic review and meta-analysis. J Investig Surg 34(3):307–317CrossRef
6.
Zurück zum Zitat Prock-Gibbs H, Pumilia CA, Meckmongkol T, Lovejoy J, Mumith A, Coathup M (2021) Incidence of osteolysis and aseptic loosening following metal-on-highly cross-linked polyethylene hip arthroplasty: a systematic review of studies with up to 15-year follow-up. J Bone Jt Surg Am 103:728–740CrossRef Prock-Gibbs H, Pumilia CA, Meckmongkol T, Lovejoy J, Mumith A, Coathup M (2021) Incidence of osteolysis and aseptic loosening following metal-on-highly cross-linked polyethylene hip arthroplasty: a systematic review of studies with up to 15-year follow-up. J Bone Jt Surg Am 103:728–740CrossRef
7.
Zurück zum Zitat Takahashi Y, Tateiwa T, Pezzotti G, Shishido T, Masaoka T, Yamamoto K (2016) Improved resistance to neck-liner impingement in second-generation highly crosslinked polyethylene—the role of vitamin E and crosslinks. J Arthroplast 31:2926–2932CrossRef Takahashi Y, Tateiwa T, Pezzotti G, Shishido T, Masaoka T, Yamamoto K (2016) Improved resistance to neck-liner impingement in second-generation highly crosslinked polyethylene—the role of vitamin E and crosslinks. J Arthroplast 31:2926–2932CrossRef
8.
Zurück zum Zitat Yamamoto K, Tateiwa T, Takahashi Y (2017) Vitamin E-stabilized highly crosslinked polyethylenes: the role and effectiveness in total hip arthroplasty. J Orthop Sci 22:384–390CrossRefPubMed Yamamoto K, Tateiwa T, Takahashi Y (2017) Vitamin E-stabilized highly crosslinked polyethylenes: the role and effectiveness in total hip arthroplasty. J Orthop Sci 22:384–390CrossRefPubMed
9.
Zurück zum Zitat Lambert B, Neut D, van der Veen HC, Bulstra SK (2019) Effects of vitamin E incorporation in polyethylene on oxidative degradation, wear rates, immune response, and infections in total joint arthroplasty: a review of the current literature. Int Orthop 43:1549–1557CrossRefPubMed Lambert B, Neut D, van der Veen HC, Bulstra SK (2019) Effects of vitamin E incorporation in polyethylene on oxidative degradation, wear rates, immune response, and infections in total joint arthroplasty: a review of the current literature. Int Orthop 43:1549–1557CrossRefPubMed
10.
Zurück zum Zitat Jameson SS, Lees D, James P, Serrano-Pedraza I, Partington PF, Muller SD, Meek RM, Reed MR (2011) Lower rates of dislocation with increased femoral head size after primary total hip replacement: a five-year analysis of NHS patients in England. J Bone Jt Surg Br 93:876–880CrossRef Jameson SS, Lees D, James P, Serrano-Pedraza I, Partington PF, Muller SD, Meek RM, Reed MR (2011) Lower rates of dislocation with increased femoral head size after primary total hip replacement: a five-year analysis of NHS patients in England. J Bone Jt Surg Br 93:876–880CrossRef
11.
Zurück zum Zitat Sariali E, Lazennec JY, Khiami F, Catonné Y (2009) Mathematical evaluation of jumping distance in total hip arthroplasty: influence of abduction angle, femoral head offset, and head diameter. Acta Orthop 80:277–282CrossRefPubMedPubMedCentral Sariali E, Lazennec JY, Khiami F, Catonné Y (2009) Mathematical evaluation of jumping distance in total hip arthroplasty: influence of abduction angle, femoral head offset, and head diameter. Acta Orthop 80:277–282CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Bartel DL, Burstein AH, Toda MD, Edwards DL (1985) The effect of conformity and plastic thickness on contact stresses in metal-backed plastic implants. J Biomech Eng 107:193–199CrossRefPubMed Bartel DL, Burstein AH, Toda MD, Edwards DL (1985) The effect of conformity and plastic thickness on contact stresses in metal-backed plastic implants. J Biomech Eng 107:193–199CrossRefPubMed
13.
Zurück zum Zitat Takahashi Y, Tateiwa T, Shishido T, Masaoka T, Kubo K, Yamamoto K (2016) Size and thickness effect on creep behavior in conventional and vitamin E-diffused highly crosslinked polyethylene for total hip arthroplasty. J Mech Behav Biomed Mater 62:399–406CrossRefPubMed Takahashi Y, Tateiwa T, Shishido T, Masaoka T, Kubo K, Yamamoto K (2016) Size and thickness effect on creep behavior in conventional and vitamin E-diffused highly crosslinked polyethylene for total hip arthroplasty. J Mech Behav Biomed Mater 62:399–406CrossRefPubMed
14.
Zurück zum Zitat Tateiwa T, Affatato S, Takahashi Y, Shishido T, Pezzotti G, Yamamoto K (2022) To what extent could the acetabular liner thickness be reduced yet remaining tribologically acceptable in metal-on-vitamin E-diffused crosslinked polyethylene hip arthroplasty? J Biomed Mater Res B Appl Biomater 110:2299–2309CrossRefPubMed Tateiwa T, Affatato S, Takahashi Y, Shishido T, Pezzotti G, Yamamoto K (2022) To what extent could the acetabular liner thickness be reduced yet remaining tribologically acceptable in metal-on-vitamin E-diffused crosslinked polyethylene hip arthroplasty? J Biomed Mater Res B Appl Biomater 110:2299–2309CrossRefPubMed
15.
Zurück zum Zitat Ong KL, Rundell S, Liepins I, Laurent R, Markel D, Kurtz SM (2009) Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press-fit implantation. J Orthop Res 27:1467–1472CrossRefPubMed Ong KL, Rundell S, Liepins I, Laurent R, Markel D, Kurtz SM (2009) Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press-fit implantation. J Orthop Res 27:1467–1472CrossRefPubMed
16.
Zurück zum Zitat Goebel P, Kluess D, Wieding J, Souffrant R, Heyer H, Sander M, Bader R (2013) The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis. J Orthop Sci 18:264–270CrossRefPubMed Goebel P, Kluess D, Wieding J, Souffrant R, Heyer H, Sander M, Bader R (2013) The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis. J Orthop Sci 18:264–270CrossRefPubMed
17.
Zurück zum Zitat Kamada K, Takahashi Y, Tateiwa T, Shishido T, Masaoka T, Pezzotti G, Yamamoto K (2020) Influence of liner offset and locking mechanism on fatigue durability in highly cross-linked polyethylene total hip prostheses. J Biomed Mater Res B Appl Biomater 108:1993–2004CrossRefPubMed Kamada K, Takahashi Y, Tateiwa T, Shishido T, Masaoka T, Pezzotti G, Yamamoto K (2020) Influence of liner offset and locking mechanism on fatigue durability in highly cross-linked polyethylene total hip prostheses. J Biomed Mater Res B Appl Biomater 108:1993–2004CrossRefPubMed
18.
Zurück zum Zitat Langlois J, Hamadouche M (2023) What have we learned from 20 years of using highly crosslinked PE in total hip arthroplasty? Orthop Traumatol Surg Res 109:103457CrossRefPubMed Langlois J, Hamadouche M (2023) What have we learned from 20 years of using highly crosslinked PE in total hip arthroplasty? Orthop Traumatol Surg Res 109:103457CrossRefPubMed
19.
Zurück zum Zitat Girard J (2015) Femoral head diameter considerations for primary total hip arthroplasty. Orthop Traumatol Surg Res 101:S25–S29CrossRefPubMed Girard J (2015) Femoral head diameter considerations for primary total hip arthroplasty. Orthop Traumatol Surg Res 101:S25–S29CrossRefPubMed
20.
Zurück zum Zitat Oonishi H, Iwaki H, Kin N, Kushitani S, Murata N, Wakitani S, Imoto K (1998) The effects of polyethylene cup thickness on wear of total hip prostheses. J Mater Sci Mater Med 9:475–478CrossRefPubMed Oonishi H, Iwaki H, Kin N, Kushitani S, Murata N, Wakitani S, Imoto K (1998) The effects of polyethylene cup thickness on wear of total hip prostheses. J Mater Sci Mater Med 9:475–478CrossRefPubMed
21.
Zurück zum Zitat Ast MP, John TK, Labbisiere A, Robador N, Valle AG (2014) Fractures of a single design of highly cross-linked polyethylene acetabular liners: an analysis of voluntary reports to the United States Food and Drug Administration. J Arthroplast 29:1231–1235CrossRef Ast MP, John TK, Labbisiere A, Robador N, Valle AG (2014) Fractures of a single design of highly cross-linked polyethylene acetabular liners: an analysis of voluntary reports to the United States Food and Drug Administration. J Arthroplast 29:1231–1235CrossRef
22.
Zurück zum Zitat Kabo JM, Gebhard JS, Loren G, Amstutz HC (1993) In vivo wear of polyethylene acetabular components. J Bone Jt Surg Br 75:254–258CrossRef Kabo JM, Gebhard JS, Loren G, Amstutz HC (1993) In vivo wear of polyethylene acetabular components. J Bone Jt Surg Br 75:254–258CrossRef
23.
Zurück zum Zitat Shen FW, Lu Z, McKellop HA (2011) Wear versus thickness and other features of 5-Mrad crosslinked UHMWPE acetabular liners. Clin Orthop Relat Res 469:395–404CrossRefPubMed Shen FW, Lu Z, McKellop HA (2011) Wear versus thickness and other features of 5-Mrad crosslinked UHMWPE acetabular liners. Clin Orthop Relat Res 469:395–404CrossRefPubMed
24.
Zurück zum Zitat Archard JF (1959) The temperature of rubbing surfaces. Wear 2:438–455CrossRef Archard JF (1959) The temperature of rubbing surfaces. Wear 2:438–455CrossRef
25.
Zurück zum Zitat Shareghi B, Johanson PE, Kärrholm J (2017) Wear of vitamin E-infused highly cross-linked polyethylene at five years. J Bone Jt Surg Am 99:1447–1452CrossRef Shareghi B, Johanson PE, Kärrholm J (2017) Wear of vitamin E-infused highly cross-linked polyethylene at five years. J Bone Jt Surg Am 99:1447–1452CrossRef
26.
Zurück zum Zitat Salemyr M, Muren O, Ahl T, Bodén H, Chammout G, Stark A, Sköldenberg O (2015) Vitamin-E diffused highly cross-linked polyethylene liner compared to standard liners in total hip arthroplasty. A randomized, controlled trial. Int Orthop 39:1499–1505CrossRefPubMed Salemyr M, Muren O, Ahl T, Bodén H, Chammout G, Stark A, Sköldenberg O (2015) Vitamin-E diffused highly cross-linked polyethylene liner compared to standard liners in total hip arthroplasty. A randomized, controlled trial. Int Orthop 39:1499–1505CrossRefPubMed
27.
Zurück zum Zitat Nebergall AK, Greene ME, Laursen MB, Nielsen PT, Malchau H, Troelsen A (2017) Vitamin E diffused highly cross-linked polyethylene in total hip arthroplasty at five years: a randomised controlled trial using radiostereometric analysis. Bone Jt J 99:577–584CrossRef Nebergall AK, Greene ME, Laursen MB, Nielsen PT, Malchau H, Troelsen A (2017) Vitamin E diffused highly cross-linked polyethylene in total hip arthroplasty at five years: a randomised controlled trial using radiostereometric analysis. Bone Jt J 99:577–584CrossRef
28.
Zurück zum Zitat Galea VP, Connelly JW, Shareghi B, Kärrholm J, Sköldenberg O, Salemyr M, Laursen MB, Muratoglu O, Bragdon C, Malchau H (2018) Evaluation of in vivo wear of vitamin E-diffused highly crosslinked polyethylene at five years: a multicentre radiostereometric analysis study. Bone Jt J 100:1592–1599CrossRef Galea VP, Connelly JW, Shareghi B, Kärrholm J, Sköldenberg O, Salemyr M, Laursen MB, Muratoglu O, Bragdon C, Malchau H (2018) Evaluation of in vivo wear of vitamin E-diffused highly crosslinked polyethylene at five years: a multicentre radiostereometric analysis study. Bone Jt J 100:1592–1599CrossRef
29.
Zurück zum Zitat Lindalen E, Thoen PS, Nordsletten L, Høvik Ø, Röhrl SM (2019) Low wear rate at 6-year follow-up of vitamin E-infused cross-linked polyethylene: a randomised trial using 32- and 36-mm heads. Hip Int 29:355–362CrossRef Lindalen E, Thoen PS, Nordsletten L, Høvik Ø, Röhrl SM (2019) Low wear rate at 6-year follow-up of vitamin E-infused cross-linked polyethylene: a randomised trial using 32- and 36-mm heads. Hip Int 29:355–362CrossRef
30.
Zurück zum Zitat Kjærgaard K, Ding M, Jensen C, Bragdon C, Malchau H, Andreasen CM, Ovesen O, Hofbauer C, Overgaard S (2020) Vitamin E-doped total hip arthroplasty liners show similar head penetration to highly cross-linked polyethylene at five years: a multi-arm randomized controlled trial. Bone Jt J 102:1303–1310CrossRef Kjærgaard K, Ding M, Jensen C, Bragdon C, Malchau H, Andreasen CM, Ovesen O, Hofbauer C, Overgaard S (2020) Vitamin E-doped total hip arthroplasty liners show similar head penetration to highly cross-linked polyethylene at five years: a multi-arm randomized controlled trial. Bone Jt J 102:1303–1310CrossRef
31.
Zurück zum Zitat Thoen PS, Nordsletten L, Pripp AH, Röhrl SM (2020) Results of a randomized controlled trial with five-year radiostereometric analysis results of vitamin E-infused highly crosslinked versus moderately crosslinked polyethylene in reverse total hip arthroplasty. Bone Jt J 102:1646–1653CrossRef Thoen PS, Nordsletten L, Pripp AH, Röhrl SM (2020) Results of a randomized controlled trial with five-year radiostereometric analysis results of vitamin E-infused highly crosslinked versus moderately crosslinked polyethylene in reverse total hip arthroplasty. Bone Jt J 102:1646–1653CrossRef
32.
Zurück zum Zitat Bergvinsson H, Zampelis V, Sundberg M, Tjörnstrand J, Flivik G (2022) Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily: a randomized, controlled RSA trial with 5-year follow-up. Acta Orthop 93:249–255CrossRefPubMedPubMedCentral Bergvinsson H, Zampelis V, Sundberg M, Tjörnstrand J, Flivik G (2022) Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily: a randomized, controlled RSA trial with 5-year follow-up. Acta Orthop 93:249–255CrossRefPubMedPubMedCentral
33.
Zurück zum Zitat Galea VP, Rojanasopondist P, Laursen M, Muratoglu OK, Malchau H, Bragdon C (2019) Evaluation of vitamin E-diffused highly crosslinked polyethylene wear and porous titanium-coated shell stability: a seven-year randomized control trial using radiostereometric analysis. Bone Jt J 101:760–767CrossRef Galea VP, Rojanasopondist P, Laursen M, Muratoglu OK, Malchau H, Bragdon C (2019) Evaluation of vitamin E-diffused highly crosslinked polyethylene wear and porous titanium-coated shell stability: a seven-year randomized control trial using radiostereometric analysis. Bone Jt J 101:760–767CrossRef
34.
Zurück zum Zitat Dumbleton JH, Manley MT, Edidin AA (2002) A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplast 17:649–661CrossRef Dumbleton JH, Manley MT, Edidin AA (2002) A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplast 17:649–661CrossRef
35.
Zurück zum Zitat Bragdon CR, Martell JM, Greene ME, Estok DM 2nd, Thanner J, Kärrholm J, Harris WH, Malchau H (2006) Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res 448:52–57CrossRefPubMed Bragdon CR, Martell JM, Greene ME, Estok DM 2nd, Thanner J, Kärrholm J, Harris WH, Malchau H (2006) Comparison of femoral head penetration using RSA and the Martell method. Clin Orthop Relat Res 448:52–57CrossRefPubMed
36.
Zurück zum Zitat Moore KD, Barrack RL, Sychterz CJ, Sawhney J, Yang AM, Engh CA (2000) The effect of weight-bearing on the radiographic measurement of the position of the femoral head after total hip arthroplasty. J Bone Jt Surg Am 82:62–69CrossRef Moore KD, Barrack RL, Sychterz CJ, Sawhney J, Yang AM, Engh CA (2000) The effect of weight-bearing on the radiographic measurement of the position of the femoral head after total hip arthroplasty. J Bone Jt Surg Am 82:62–69CrossRef
Metadaten
Titel
In vivo creep and wear performance of vitamin-E-diffused highly crosslinked polyethylene in total hip arthroplasty
verfasst von
Tsunehiro Matsuoka
Yasuhito Takahashi
Tsunehito Ishida
Toshiyuki Tateiwa
Takaaki Shishido
Kengo Yamamoto
Publikationsdatum
12.07.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 12/2023
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-023-04972-1

Weitere Artikel der Ausgabe 12/2023

Archives of Orthopaedic and Trauma Surgery 12/2023 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.