Erschienen in: BMC Pediatrics 1/2020

Open Access 01.12.2020 | Research article

Diagnostic test accuracy of new generation tympanic thermometry in children under different cutoffs: a systematic review and meta-analysis

verfasst von: Dan Shi, Li-Yuan Zhang, Hai-Xia Li

Erschienen in: BMC Pediatrics | Ausgabe 1/2020

Abstract

Background

The infrared tympanic thermometer (IRTT) is a popular method for temperature screening in children, but it has been debated for the low accuracy and reproducibility compared with other measurements. This study was aimed to identify and quantify studies reporting the diagnostic accuracy of the new generation IRTT in children and to compare the sensitivity and specificity of IRTT under different cutoffs and give the optimal cutoff.

Methods

Articles were derived from a systematic search in PubMed, Web of Science Core Collection, and Embase, and were assessed for internal validity by the Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2). The figure of risk of bias was created by Review Manager 5.3 and data were synthesized by MetaDisc 1.4.

Results

Twelve diagnostic studies, involving 4639 pediatric patients, were included. The cut-offs varied from 37.0 °C to 38.0 °C among these studies. The cut-off 37.8 °C was with the highest sROC AUC (0.97) and Youden Index (0.83) and was deemed to be the optimal cutoff.

Conclusion

The optimal cutoff for infrared tympanic thermometers is 37.8 °C. New Generation Tympanic Thermometry is with high diagnostic accuracy in pediatric patients and can be an alternative for fever screening in children.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
IRTT
Infrared tympanic thermometer
FP
The false Positive
FN
The false Negative
NPV
Negative predictive value
PPV
Positive predictive value
The PRISMA-DTA Statement
the Preferred Reporting Items for Systematic Reviews and Meta-Analyses of Diagnostic Test Accuracy Studies
QUADAS-2
The Quality Assessment of Diagnostic Accuracy Studies-2
TP
The True Positive
TN
The True Negative

Background

Body temperature measurement is a routine in the management of sick children for both parents and healthcare providers [1, 2]. An accurate diagnosis of fever is crucial in clinical practice [3, 4] and an inaccurate one could lead to serious complications and improper medical decisions [3, 5]. Core temperature is the gold standard for temperature measurement [3]. However, core temperature measurements, such as pulmonary artery and lower esophagus measurement, are invasive and require specialized equipment, therefore, are unpractical for daily clinical practice [3, 6]. Ideally, body temperature measurement should be noninvasive, accurate, pain-free, cost-effective and time-efficient [3, 7, 8].
Traditionally, non-invasive methods of body temperature measurement include rectal temperature, oral temperature and axillary temperature. Among these methods, rectal thermometry has been the most reliable for measuring body temperature in children and is considered clinically to be the best estimation of the core temperature [9]. However, it is time-consuming and requires certain level of practice [5, 10]. Furthermore, it may cause emotional distress, and -although very rare- brings possible complications such as perforation or transmission of micro-organisms [5, 10]. And therefore infants, health workers and parents more or less express reluctance to perform it [3].
The forehead skin thermometer (FST) and infrared tympanic thermometer (IRTT) are popular alternatives for the traditional measures. The FST uses a sensor probe to measure the amount of infrared heat produced by the temporal arteries [8]. The IRTT detects the radiation of tympanic membrane and the ear canal, which share the blood supply with the hypothalamus, the thermoregulatory center of the human body [11, 12]. Both these two methods are safe, easy to use, comfortable and quick. But compared to the FST, the IRTT is more consistent with rectal temperature and is more convincing [3, 8, 13]. Using the aural temperature is less traumatic and allows a faster triage [14], but it has been debated for the low accuracy and reproducibility compared with other measurements [1, 1418]. Over the past years, however, the IRTT have been developed and updated, and some older versions have been obsolete. The new generation IRTT used various brand-specific ways to enhance accuracy, for example, improvements of geometry and algorithms, a wider angle measurement, displaying temperature on multiple samples and equipping with a heat probe [11, 19]. Synthesizing studies applying obsolete IRTT with the new ones is unreasonable and may underestimate the IRTT test accuracy. Furthermore, the cutoffs of the IRTT used in fever detection are diverse, and the optimal cut-off has no consensus. The cutoff means a temperature threshold that divides pediatric patients into fever and non-fever, and the diagnostic accuracy of IRTT various under different cutoffs [3, 13, 20, 21]. It is inappropriate to synthesize studies applying different cutoffs and the results are unreliable.
The aims of this systematic review were (1) to identify and quantify studies reporting the diagnostic accuracy of the new generation of the IRTT in children (By new generation, we meant the IRTT that were still in production and on sale according to the official websites of the manufacturers as we started our study); (2) to compare the sensitivity and specificity under different cutoffs of the IRTT and give the optimal cutoff.

Methods

Search strategies

The conduct of this systematic review and meta-analysis was based on the Test Accuracy Working Group of the Cochrane Collaboration and the Preferred Reporting Items for Systematic Reviews and Meta-Analyses of Diagnostic Test Accuracy Studies statement (The PRISMA-DTA Statement) guidelines [22, 23]. A systematic literature search of multiple electronic databases (PubMed, Web of Science Core Collection, EMBASE) was conducted by two trained reviewers (D.S. and LY.Z.) independently from inception to February 2nd, 2019. The following search terms ((tympanic thermometer OR ear thermometer OR infrared thermometry OR ear thermometry OR tympanic scan OR tympanic temperature OR ear temperature OR infrared thermometer OR ear thermometer)) AND (pediatric OR child OR kid OR newborn OR baby OR infant OR toddler) in All Fields (PubMed, EMBASE) or Topic (Web of Science Core Collection) were used. The languages were restricted to English and species were restricted to humans. The bibliographies of included studies were also searched to identify additional studies.

Study selection

Observational studies, detecting fever by aural and rectal thermometers, were deemed acceptable. Inclusion criterion included (1) studies recruiting pediatric subjects (age < 18 years), (2) diagnostic test accuracy studies, (3) studies detecting fever by new generation IRTT, and (4) studies using rectal thermometers as the reference standard. Exclusion criterion included (1) studies unrelated to the accuracy of IRTT, (2) reviews, proceedings papers, meeting abstracts, letters, notes and editorial materials, and (3) studies lacking essential data.
Two reviewers (D.S. and LY.Z.) independently reviewed the titles and abstracts of these studies. Papers deemed to match the predefined inclusion criteria or without consensus were reviewed in full text. Disagreements were resolved through discussions and scientific consultations.

Quality assessment and data extraction

We adopted the Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2, [24] for quality assessment and used Review Manager 5.3 for creating the figures of risk of bias and applicability concerns [25]. Two independent reviewers (D.S. and LY.Z.) assessed the methodological quality of the included studies independently and disagreements were also resolved through discussions and scientific consultations.
The following data were extracted by two independent reviewers (D.S. and LY.Z) from the included studies: (1) descriptive aspects: primary author, year of publication, country, setting, age, types of tympanic thermometer and reference standard; (2) statistical aspects: the size, number of observations, the cut-off of tympanic thermometer, the True Positive (TP), the False Negative (FN), the False Positive (FP) and the True Negative (TN), sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV).

Statistical analysis

Meta-analyses of TP, FN, FP and TN were performed to compare the test accuracy between tympanic temperature and the gold standard (rectal temperature) by MetaDiSc 1.4 [26]. Threshold analysis was conducted to evaluate the threshold effect [27]. The inconsistency index (I2) test was used to estimate heterogeneity between studies and I2 > 75% was considered to be with high heterogeneity [28]. Data were synthesized by using the random-effects model which was recommended in pooled estimates of diagnostic meta-analyses [29]. The area under the curve (AUC), Youden index and index Q* were used to measure test accuracy [3032].

Results

Selection process

Initially, 611, 468 and 276 articles were retrieved from PubMed, Web of Science Core Collection and EMBASE respectively. Secondly, 332 duplicates were removed. Thirdly, the titles and abstracts of the remaining 1023 articles were examined and 975 articles were excluded for diverse reasons. Finally, 11 articles were selected after the full text review and 1 article [33] was added by reviewing references. The process and outcome of the literature selection are presented in detail in Fig. 1.

Risk of bias and applicability concerns in included studies

Figure 2 and Fig. 3 showed the risk of bias and applicability concerns in different domains. Among these 12 included articles, 4 had a high risk of bias on “flow and timing”, “patient selection”, “index test”, and “reference standard”, indicting the quality Methodological quality of included studies was moderate. Eight out of twelve studies had low applicability concerns in all domains and the applicability concerns was low.

Characteristics of selected studies

Twelve included studies were published from 2010 to 2018. All these studies applied the tympanic thermometer and set the rectal thermometer as reference standard. The descriptive and statistical characteristics of the 12 studies were presented in Table 1 and Table 2 respectively.
Table 1
Descriptive characteristics of including studies
Studies
Year
Country
Setting
Age
Tympanic thermometer
Reference standard
Mogensen et al [13]
2018
Denmark
pediatric emergency department
0-18y
Braun Welch Allyn Pro 4000 Thermoscan
Rectal
Paramita et al [33]
2017
Indonesia
Pediatric outpatient clinic/ pediatric emergency department/ inpatient pediatrics ward
6 m-5y
OMRON Gentle Temp 510
Rectal
Chatproedprai et al [3]
2016
Thailand
Pediatric outpatient clinic
0-2y
Microlife IR1DE1–1
Rectal
Acikgoz et al [30]
2016
Turkey
pediatric emergency department
3 m-3y
Genius™ 2
Rectal
Allegaert et al [5]
2014
Belgium
Pediatric wards
2 m-17y
Genius™ 2
Rectal
Teller et al [34]
2014
Swizerland
Pediatric practice
1 m-2y
Braun Thermoscan 6022™
Rectal
Hamilton et al [15]
2014
America
The emergency department and the overflow patient treatment areas
0-18y
Braun Welch Allyn Pro 4000 Thermoscan
Rectal
Batra et al [29]
2013
India
The pediatric emergency room
2-12y
Equinox infrared ear thermometer (EQ. ET 99)
Rectal
Duru et al. [31]
2012
Nigeria
The neonatal wards
6.63 ± 6.98d
Braun IRT 4520 Thermoscan
Rectal
Edelu et al [35]
2011
Nigeria
Pediatric outpatient clinic/ pediatric emergency department
0-5y
OMRON instant ear thermometer model MC-509 N
Rectal
Paes et al [8]
2010
Netherlands
The pediatric ward
0-18y
The FirstTemp Genius tympanic thermometer 3000A
Rectal
Oyakhirome et al [32]
2010
Gabon
The outpatient department
0-10y
Braun 6022 Thermoscan
Rectal
Table 2
Statistical characteristics of including studies
Studies
Sample
Cut-off
TP
FP
FN
TN
Se
Sp
PPV
NPV
Mogensen et al [13]
995
37.8
372
76
20
527
95
87
83
96
 
38.0
350
36
43
566
89
94
91
93
Paramita et al [33]
90
37.4
65
14
3
8
96
36
82
73
 
37.5
64
11
4
11
94
50
85
73
 
37.6
63
11
5
11
93
50
85
69
 
37.7
62
9
6
13
91
60
87
68
 
37.8
60
6
8
16
88
73
91
66
Chatproedprai et al [3]
312
37.0
181
17
22
92
89
84
91
81
 
37.6
126
1
77
108
62
99
99
58
Acikgoz et al [30]
354
37.25
163
22
33
136
83
86
88
80
Allegaert et al [5]
294
38.0
5
0
17
272
22
100
100
94
Teller et al [34]
254
38.0
72
4
28
150
72
97
95
84
 
37.6
93
25
7
129
93
84
79
95
Hamilton et al [15]
205
38.0
87
8
6
104
94
93
92
95
Batra et al [29]
100
38.0
49
1
1
49
98
98
98
98
Duru et al [31]
300
37.5
34
3
11
252
76
99
93
97
Edelu et al [35]
710
37.8
316
10
39
345
89
97
97
90
90
37.6
33
4
12
41
73
91
89
77
Paes et al [8]
100
38.0
20
2
5
73
80
97
91
94
Oyakhirome et al [32]
835
38.0
337
19
112
357
75
95
94
76

Accuracy of tympanic thermometry in children under different cut-offs

The 12 studies involved 4639 children. The cut-off points were various. Among the included articles, 7 [5, 8, 18, 3336] studies set the optimal cut-off and the other 5 [3, 13, 14, 20, 21] studies analyzed the diagnostic test accuracy of tympanic thermometry under different cut-offs. The range of the cut-off point was from 37.0 °C to 38.0 °C. Studies had data under same cut-off were synthesized.

Accuracy under the cut-off of 37.0 °C

There was only one study [3] reported diagnostic test accuracy under the cut-off 37.0 °C. In this study, for ear temperature (37.0 °C), sensitivity, specificity, PPV, and NPV were 0.89, 0.84, 0.91, and 0.81 respectively.

Accuracy under the cut-off of 37.25 °C

Only one study [34] gave the optimal cut-off 37.25 °C and sensitivity, specificity, PPV, and NPV were 0.83, 0.86, 0.88, and 0.80 respectively.

Accuracy under the cut-off of 37.4 °C

There was only one study [20] reported diagnostic test accuracy under the cut-off 37.4 °C. In this study, for ear temperature (37.4 °C), sensitivity, specificity, PPV, and NPV were 0.96, 0.36, 0.82, and 0.73 respectively.

Accuracy under the cut-off of 37.5 °C

The cut-off 37.5 °C was used in 2 studies [20, 35] and a total of 390 pediatric patients were involved. The pooled sensitivity was 0.87 (95% CI 0.79–0.92) and heterogeneity between the articles was high: 87.5% (X2 = 8.02, P < 0.05). The pooled specificity was 0.95 (95% CI 0.92–0.97) and heterogeneity between the articles was high: 97.9% (X2 = 47.74, P < 0.05).

Accuracy under the cut-off of 37.6 °C

The cut-off 37.6 °C was used in 4 studies [3, 13, 20, 21] and a total of 746 pediatric patients were involved. Spearman’s correlation coefficient of sensitivity and specificity was 0.089 (P = .638) and the ROC plane showed no curvilinear trend, suggesting that there was no heterogeneity from threshold effect. The pooled sensitivity was 0.76 (95% CI 0.71–0.80) and heterogeneity between the articles was high: 94.3% (X2 = 53.04, P < 0.05). The pooled specificity was 0.88 (95% CI 0.84–0.91) and heterogeneity between the articles was high: 92.9% (X2 = 42.22, P < 0.05) (Fig. 4). The sROC AUC was 0.93 (SE = 0.02) while Q* value was 0.86 (SE = 0.03).

Accuracy under the cut-off of 37.7 °C

There was only one study [20] reported diagnostic test accuracy under the cut-off 37.7 °C. In this study, for ear temperature (37.7 °C), sensitivity, specificity, PPV, and NPV were 0.91, 0.60, 0.87, and 0.68 respectively.

Accuracy under the cut-off of 37.8 °C

The cut-off 37.8 °C was used in 3 studies [14, 20, 21] and a total of 1795 pediatric patients were involved. The threshold analysis (r = − 0.050, P = .667) and the ROC plane (Figure) suggested that there was no heterogeneity from threshold effect. The pooled sensitivity was 0.92 (95% CI 0.90–0.94) and heterogeneity between the articles was high: 80.1% (X2 = 10.07, P < 0.05). The pooled specificity was 0.91 (95% CI 0.89–0.92) and heterogeneity between the articles was high: 94.5% (X2 = 36.68, P < 0.05) (Fig. 5). The sROC AUC was 0.97 (SE = 0.02) while Q* value was 0.91 (SE = 0.03).

Accuracy under the cut-off of 38.0 °C

The cut-off 38.0 °C was used in 7 studies [5, 8, 13, 14, 18, 33, 36] and a total of 2783 pediatric patients were involved. The threshold analysis (r = 0.429, P = 0.337) and the ROC plane suggested that there was no heterogeneity from threshold effect. The pooled sensitivity was 0.81 (95% CI 0.79–0.84) and heterogeneity between the articles was high: 93.7% (X2 = 94.51, P < 0.05). The pooled specificity was 0.96 (95% CI 0.95–0.97) and heterogeneity between the articles was high: 81.6% (X2 = 32.56, P < 0.05) (Fig. 6). The sROC AUC was 0.97 (SE = 0.01) while Q* value was 0.92 (SE = 0.01).
The diagnostic test accuracy of the tympanic thermometry under different Cut-offs in the detection of pediatric fever is summarized in Table 3. The cut-off 37.8 is with the highest sROC AUC and Youden Index and is deemed to be the optimal cutoff.
Table 3
Accuracy of tympanic thermometry under different cutoffs in children
Cut-off(°C)
N
Sen
Sp
Youden Index
SROC-AUC
37.0
312
89
84
0.73
N
37.25
354
83
86
0.69
N
37.4
90
96
36
0.32
N
37.5
390
87
95
0.82
N
37.6
746
76
88
0.64
0.93
37.7
90
91
60
0.51
N
37.8
1795
92
91
0.83
0.97
38.0
2578
80
96
0.76
0.97

Discussion

We conducted this study to assess the discriminant validity of the new generation IRTT for detecting pediatric fever determined by rectal thermometry and to find the optimal cutoff. Twelve studies, including 4639 children, were included. The results indicated that IRTT was a good alternative for rectal thermometry in pediatric patients, and the optimal cut-off of ear temperature for screening fever in children was 37.8 °C. Under this cut-off, pooled sensitivity was 0.92 (95% CI 0.90–0.94), pooled specificity was 0.91 (95% CI 0.89–0.92), sROC AUC was 0.97 (SE = 0.02) and Q* value was 0.91 (SE = 0.03).
One major strength of this study was that it estimated the test accuracy of new generation IRTT. Although the IRTT may provide a good alternative for traditional measurements, it has been debated for the low reproducibility. However, since the ear thermometer came out, it has been constantly updated and upgraded. Some techniques have been used to improve the test accuracy, such as the Braun Welch Allyn Pro 4000 Thermoscan, where a heating element in the sensor heats the probe tip to just below normal body temperature to avoid cooling the ear canal [19]. And the improvements of geometry and algorithms have been developed to ensure that the displayed result reflects the tympanic temperature accurately [11]. Hence, the newer versions of tympanic thermometers might meet the clinicians’ requested improvements of repeatability in noninvasive temperature assessments. By new generation, we mean the IRTT that were still in production and on sale according to the official websites of the manufacturers as we started our study. We included the tympanic thermometers under use and excluded the outdated ones so that the results could provide a reference for current clinical practice.
Another strength of this study was that it estimated the test accuracy of new generation IRTT under different cutoffs. The synthesis of data under different cutoffs may underestimated the test accuracy of IRTT, because the diagnostic accuracy of IRTT varied under different cutoffs [3, 13, 20, 21]. The cutoffs of IRTT ranged from 37.0 °C to 38 °C among these 12 included studies. After the synthesis of three studies, including 1795 children, we found the optimal cut-off of tympanic thermometry is 37.8 °C. And under this cutoff, the pooled sensitivity was 0.92 (95% CI 0.90–0.94), pooled specificity was 0.91 (95% CI 0.89–0.92), sROC AUC was 0.97 (SE = 0.02) and Q* value was 0.91 (SE = 0.03).
The diagnostic accuracy in this study under the optimal cutoff was far higher than a former systematic review [27], in which pooled sensitivity was 0.70 (95% CI 0.68–0.72), pooled specificity was 0.86 (95% CI 0.85–0.88), sROC AUC was 0.94, and Q* value was 0.87. Excluding articles applying obsolete tympanic thermometers and analyzing diagnostic test accuracy under different cut-offs may be the major reasons for this gap.
The 12 included studies are with high homogeneity, because they have the same study type, study population, reference standard and et al. And data were synthesized by using the random-effects model. What should be underlined is that the heterogeneity between the articles is very high, from 81.6 to 94.5%. The study population of included studies are all children, who age from 0 to 18-year-old. But the age groups are various, for example, Duru et al. [35] admitted neonates whose mean age is 6.63 ± 6.98 days, while Allegaert et al. [5] enrolled children with a median age of 3.2 years (range 0.02 years to 17 years). The variation of age groups may be the major contribution to the high heterogeneity and further studies focusing on different age groups are needed.
Although the results of our study can provide an important reference for subsequent researches and clinical applications, there are two limitations in our present study. We performed different sub-group meta-analyses based on the different cut-offs used. Unfortunately, in many of these analyses a limited number of studies are included. We concluded that 37.8 °C was the optimal cut-off just based on three studies, which seemed unconvincing. But considering that 1795 subjects were included for analysis under the cut-off 37.8 °C, the conclusion was much more convincing.
According to the findings, ear canal temperature can be confidently implemented as a screening measure in the pediatric fever detection. This application of IRTT would effectively decrease the number of children who require the rectal temperature method for fever detection [7]. However, there are some situations, such as uncertain diagnosis [7], during exercise [37, 38], change of environmental temperatures [39], that tympanic temperature should not be used as a surrogate for rectal temperature.

Conclusion

Tympanic thermometry has a high diagnostic accuracy and is a good alternative for temperature screening in pediatric patients. The optimal cut-off of ear temperature for screening fever in children is 37.8 °C. Tympanic thermometry may not be an alternative for rectal temperature after intense exercise or exertion heat stroke.

Acknowledgements

Not applicable.
Not applicable.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Dogan HH, Sezer RG, Kirkgoz T, Bozaykut A. Comparison of axillary and tympanic temperature measurements in children diagnosed with acute otitis media. Int J Pediatr. 2016;2016:1729218.PubMedPubMedCentralCrossRef Dogan HH, Sezer RG, Kirkgoz T, Bozaykut A. Comparison of axillary and tympanic temperature measurements in children diagnosed with acute otitis media. Int J Pediatr. 2016;2016:1729218.PubMedPubMedCentralCrossRef
2.
Zurück zum Zitat Barton SJ, Gaffney R, Chase T, Rayens MK, Piyabanditkul L. Pediatric temperature measurement and child/parent/nurse preference using three temperature measurement instruments. J Pediatr Nurs. 2003;18(5):314–20.PubMedCrossRef Barton SJ, Gaffney R, Chase T, Rayens MK, Piyabanditkul L. Pediatric temperature measurement and child/parent/nurse preference using three temperature measurement instruments. J Pediatr Nurs. 2003;18(5):314–20.PubMedCrossRef
3.
Zurück zum Zitat Chatproedprai S, Heamawatanachai K, Tempark T, Wananukul S. A comparative study of 3 different methods of temperature measurement in children. J Med Assoc Thail. 2016;99(2):142–9. Chatproedprai S, Heamawatanachai K, Tempark T, Wananukul S. A comparative study of 3 different methods of temperature measurement in children. J Med Assoc Thail. 2016;99(2):142–9.
4.
Zurück zum Zitat Davis T. NICE guideline: feverish illness in children—assessment and initial management in children younger than 5 years. Arch Dis Child Educ Pract Ed. 2013;98:232–5.PubMedCrossRef Davis T. NICE guideline: feverish illness in children—assessment and initial management in children younger than 5 years. Arch Dis Child Educ Pract Ed. 2013;98:232–5.PubMedCrossRef
5.
Zurück zum Zitat Allegaert K, Casteels K, van Gorp I, Bogaert G. Tympanic, infrared skin, and temporal artery scan thermometers compared with rectal measurement in children: a real-life assessment. Curr Ther Res Clin Exp. 2014;76:34–8.PubMedPubMedCentralCrossRef Allegaert K, Casteels K, van Gorp I, Bogaert G. Tympanic, infrared skin, and temporal artery scan thermometers compared with rectal measurement in children: a real-life assessment. Curr Ther Res Clin Exp. 2014;76:34–8.PubMedPubMedCentralCrossRef
6.
Zurück zum Zitat Abdulkadir MB, Johnson B. A comparative study of rectal tympanic and axillary thermometry in febrile children under 5 years of age in Nigeria. Paediatr Int Child Health. 2013;33(3):165–9.PubMedCrossRef Abdulkadir MB, Johnson B. A comparative study of rectal tympanic and axillary thermometry in febrile children under 5 years of age in Nigeria. Paediatr Int Child Health. 2013;33(3):165–9.PubMedCrossRef
7.
Zurück zum Zitat Chiappini E, Principi N, Longhi R, Tovo PA, Becherucci P, Bonsignori F, Esposito S, Festini F, Galli L, Lucchesi B, et al. Management of fever in children: summary of the Italian pediatric society guidelines. Clin Ther. 2009;31(8):1826–43.PubMedCrossRef Chiappini E, Principi N, Longhi R, Tovo PA, Becherucci P, Bonsignori F, Esposito S, Festini F, Galli L, Lucchesi B, et al. Management of fever in children: summary of the Italian pediatric society guidelines. Clin Ther. 2009;31(8):1826–43.PubMedCrossRef
8.
Zurück zum Zitat Paes BF, Vermeulen K, Brohet RM, van der Ploeg T, de Winter JP. Accuracy of tympanic and infrared skin thermometers in children. Arch Dis Child. 2010;95(12):974–8.PubMedCrossRef Paes BF, Vermeulen K, Brohet RM, van der Ploeg T, de Winter JP. Accuracy of tympanic and infrared skin thermometers in children. Arch Dis Child. 2010;95(12):974–8.PubMedCrossRef
9.
Zurück zum Zitat Dinarello CA, Wolff SM. Principles and practice of infectious diseases. 3rd ed. New York: Churchill Livingstone; 1990. Dinarello CA, Wolff SM. Principles and practice of infectious diseases. 3rd ed. New York: Churchill Livingstone; 1990.
10.
Zurück zum Zitat Apa H, Gozmen S, Bayram N, Catkoglu A, Devrim F, Karaarslan U, Gunay I, Unal N, Devrim I. Clinical accuracy of tympanic thermometer and noncontact infrared skin thermometer in pediatric practice: an alternative for axillary digital thermometer. Pediatr Emerg Care. 2013;29(9):992–7.PubMedCrossRef Apa H, Gozmen S, Bayram N, Catkoglu A, Devrim F, Karaarslan U, Gunay I, Unal N, Devrim I. Clinical accuracy of tympanic thermometer and noncontact infrared skin thermometer in pediatric practice: an alternative for axillary digital thermometer. Pediatr Emerg Care. 2013;29(9):992–7.PubMedCrossRef
11.
Zurück zum Zitat Aadal L, Fog L, Pedersen AR. Tympanic ear thermometer assessment of body temperature among patients with cognitive disturbances. An acceptable and ethically desirable alternative? Scand J Caring Sci. 2016;30(4):766–73.PubMedCrossRef Aadal L, Fog L, Pedersen AR. Tympanic ear thermometer assessment of body temperature among patients with cognitive disturbances. An acceptable and ethically desirable alternative? Scand J Caring Sci. 2016;30(4):766–73.PubMedCrossRef
12.
Zurück zum Zitat Denis Leduc SW. Val-d’Or. Canadian Paediatric Society: Community Paediatrics Committee. Temperature Measurements in Paediatrics. Ontario; 2007. Denis Leduc SW. Val-d’Or. Canadian Paediatric Society: Community Paediatrics Committee. Temperature Measurements in Paediatrics. Ontario; 2007.
13.
Zurück zum Zitat Teller J, Ragazzi M, Simonetti GD, Lava SA. Accuracy of tympanic and forehead thermometers in private paediatric practice. Acta Paediatr. 2014;103(2):e80–3.PubMedCrossRef Teller J, Ragazzi M, Simonetti GD, Lava SA. Accuracy of tympanic and forehead thermometers in private paediatric practice. Acta Paediatr. 2014;103(2):e80–3.PubMedCrossRef
14.
Zurück zum Zitat Mogensen CB, Wittenhoff L, Fruerhoj G, Hansen S. Forehead or ear temperature measurement cannot replace rectal measurements, except for screening purposes. BMC Pediatr. 2018;18(1):15.PubMedPubMedCentralCrossRef Mogensen CB, Wittenhoff L, Fruerhoj G, Hansen S. Forehead or ear temperature measurement cannot replace rectal measurements, except for screening purposes. BMC Pediatr. 2018;18(1):15.PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat Robinson JL, Seal RF, Spady DW, Joffres MR. Comparison of esophageal, rectal, axillary, bladder, tympanic, and pulmonary artery temperatures in children. J Pediatr. 1998;133(4):553–6.PubMedCrossRef Robinson JL, Seal RF, Spady DW, Joffres MR. Comparison of esophageal, rectal, axillary, bladder, tympanic, and pulmonary artery temperatures in children. J Pediatr. 1998;133(4):553–6.PubMedCrossRef
16.
Zurück zum Zitat Hamilton PA, Marcos LS, Secic M. Performance of infrared ear and forehead thermometers: a comparative study in 205 febrile and afebrile children. J Clin Nurs. 2013;22(17–18):2509–18.PubMedCrossRef Hamilton PA, Marcos LS, Secic M. Performance of infrared ear and forehead thermometers: a comparative study in 205 febrile and afebrile children. J Clin Nurs. 2013;22(17–18):2509–18.PubMedCrossRef
17.
Zurück zum Zitat Akinyinka OO, Omokhodion SI, Olawuyi JF, Olumese PE, Brown BJ. Tympanic thermometry in Nigerian children. Ann Trop Paediatr. 2001;21(2):169–74.PubMedCrossRef Akinyinka OO, Omokhodion SI, Olawuyi JF, Olumese PE, Brown BJ. Tympanic thermometry in Nigerian children. Ann Trop Paediatr. 2001;21(2):169–74.PubMedCrossRef
18.
Zurück zum Zitat Sund-Levander M, Grodzinsky E. Time for a change to assess and evaluate body temperature in clinical practice. Int J Nurs Pract. 2009;15(4):241–9.PubMedCrossRef Sund-Levander M, Grodzinsky E. Time for a change to assess and evaluate body temperature in clinical practice. Int J Nurs Pract. 2009;15(4):241–9.PubMedCrossRef
19.
Zurück zum Zitat Haugan B, Langerud AK, Kalvoy H, Froslie KF, Riise E, Kapstad H. Can we trust the new generation of infrared tympanic thermometers in clinical practice? J Clin Nurs. 2013;22(5–6):698–709.PubMed Haugan B, Langerud AK, Kalvoy H, Froslie KF, Riise E, Kapstad H. Can we trust the new generation of infrared tympanic thermometers in clinical practice? J Clin Nurs. 2013;22(5–6):698–709.PubMed
20.
Zurück zum Zitat Paramita T, Karyanti MR. Soedjatmiko, Hendarto a, Suyoko D, Latief a. a comparison of axillary and tympanic membrane to rectal temperatures in children. PAEDIATRICA INDONESIANA. 2017;57(1):47–51.CrossRef Paramita T, Karyanti MR. Soedjatmiko, Hendarto a, Suyoko D, Latief a. a comparison of axillary and tympanic membrane to rectal temperatures in children. PAEDIATRICA INDONESIANA. 2017;57(1):47–51.CrossRef
21.
Zurück zum Zitat Edelu BO, Ojinnaka NC, Ikefuna AN. Fever detection in under 5 children in a tertiary health facility using the infrared tympanic thermometer in the oral mode. Ital J Pediatr. 2011;37:8.PubMedPubMedCentralCrossRef Edelu BO, Ojinnaka NC, Ikefuna AN. Fever detection in under 5 children in a tertiary health facility using the infrared tympanic thermometer in the oral mode. Ital J Pediatr. 2011;37:8.PubMedPubMedCentralCrossRef
22.
23.
Zurück zum Zitat McInnes MDF, Moher D, Thombs BD, McGrath TA, Bossuyt PM, Clifford T, Cohen JF, Deeks JJ, Gatsonis C, Hooft L, et al. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies. JAMA. 2018;319(4):388.PubMedCrossRef McInnes MDF, Moher D, Thombs BD, McGrath TA, Bossuyt PM, Clifford T, Cohen JF, Deeks JJ, Gatsonis C, Hooft L, et al. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies. JAMA. 2018;319(4):388.PubMedCrossRef
24.
Zurück zum Zitat Greiner M, Pfeiffer D, Smith RD. Principles and practical application of the receiver-operating characteristic analysis for diagnostic tests. PREV VET MED. 2000;45(1–2):23–41.PubMedCrossRef Greiner M, Pfeiffer D, Smith RD. Principles and practical application of the receiver-operating characteristic analysis for diagnostic tests. PREV VET MED. 2000;45(1–2):23–41.PubMedCrossRef
25.
Zurück zum Zitat The Nordic Cochrane Centre, The Cochrane Collaboration. Review Manager (RevMan) [Computer program]. Version 5.3. Copenhagen: The Nordic Cochrane Centre, The Cochrane Collaboration; 2014. The Nordic Cochrane Centre, The Cochrane Collaboration. Review Manager (RevMan) [Computer program]. Version 5.3. Copenhagen: The Nordic Cochrane Centre, The Cochrane Collaboration; 2014.
26.
Zurück zum Zitat Zamora J, Abraira V, Muriel A, Khan K, Coomarasamy A. Meta-DiSc: a software for meta-analysis of test accuracy data. BMC Med Res Methodol. 2006;6:31.PubMedPubMedCentralCrossRef Zamora J, Abraira V, Muriel A, Khan K, Coomarasamy A. Meta-DiSc: a software for meta-analysis of test accuracy data. BMC Med Res Methodol. 2006;6:31.PubMedPubMedCentralCrossRef
27.
Zurück zum Zitat Zhen C, Xia Z, Ya Jun Z, Long L, Jian S, Gui Ju C, Long L. Accuracy of infrared tympanic thermometry used in the diagnosis of fever in children. CLIN PEDIATR. 2014;54(2):114–26.CrossRef Zhen C, Xia Z, Ya Jun Z, Long L, Jian S, Gui Ju C, Long L. Accuracy of infrared tympanic thermometry used in the diagnosis of fever in children. CLIN PEDIATR. 2014;54(2):114–26.CrossRef
28.
Zurück zum Zitat Higgins JP, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med. 2002;21(11):1539–58.PubMedCrossRef Higgins JP, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med. 2002;21(11):1539–58.PubMedCrossRef
30.
Zurück zum Zitat Samawi HM, Yin J, Rochani H, Panchal V. Notes on the overlap measure as an alternative to the Youden index: how are they related? Stat Med. 2017;36(26):4230–40.PubMedCrossRef Samawi HM, Yin J, Rochani H, Panchal V. Notes on the overlap measure as an alternative to the Youden index: how are they related? Stat Med. 2017;36(26):4230–40.PubMedCrossRef
31.
Zurück zum Zitat Walter SD. Properties of the summary receiver operating characteristic (SROC) curve for diagnostic test data. Stat Med. 2002;21(9):1237–56.PubMedCrossRef Walter SD. Properties of the summary receiver operating characteristic (SROC) curve for diagnostic test data. Stat Med. 2002;21(9):1237–56.PubMedCrossRef
32.
Zurück zum Zitat Shi D, Chen X, Li Z. Diagnostic test accuracy of the Montreal cognitive assessment in the detection of post-stroke cognitive impairment under different stages and cutoffs: a systematic review and meta-analysis. Neurol Sci. 2018;39(4):705–16.PubMedCrossRef Shi D, Chen X, Li Z. Diagnostic test accuracy of the Montreal cognitive assessment in the detection of post-stroke cognitive impairment under different stages and cutoffs: a systematic review and meta-analysis. Neurol Sci. 2018;39(4):705–16.PubMedCrossRef
33.
Zurück zum Zitat Batra P, Goyal S. Comparison of rectal, axillary, tympanic, and temporal artery thermometry in the pediatric emergency room. Pediatr Emerg Care. 2013;29(1):63–6.PubMedCrossRef Batra P, Goyal S. Comparison of rectal, axillary, tympanic, and temporal artery thermometry in the pediatric emergency room. Pediatr Emerg Care. 2013;29(1):63–6.PubMedCrossRef
34.
Zurück zum Zitat Acikgoz M, Guzel A, Mura N, Karli A, Sezgin U. Reliability of the infrared and chemical dot temperature measurement methods in the children admitted in the pediatric emergency unit: a prospective study. KUWAIT MED J. 2016;48(2):105–10. Acikgoz M, Guzel A, Mura N, Karli A, Sezgin U. Reliability of the infrared and chemical dot temperature measurement methods in the children admitted in the pediatric emergency unit: a prospective study. KUWAIT MED J. 2016;48(2):105–10.
35.
36.
Zurück zum Zitat Oyakhirome S, Profanter K, Kremsner PG. Assessment of fever in African children: implication for malaria trials. AM J TROP MED HYG. 2010;82(2):215–8.PubMedPubMedCentralCrossRef Oyakhirome S, Profanter K, Kremsner PG. Assessment of fever in African children: implication for malaria trials. AM J TROP MED HYG. 2010;82(2):215–8.PubMedPubMedCentralCrossRef
37.
Zurück zum Zitat Huggins R, Glaviano N, Negishi N, Casa DJ, Hertel J. Comparison of rectal and aural Core body temperature thermometry in Hyperthermic, exercising individuals: a meta-analysis. J Athl Train. 2012;47(3):329–38.PubMedPubMedCentralCrossRef Huggins R, Glaviano N, Negishi N, Casa DJ, Hertel J. Comparison of rectal and aural Core body temperature thermometry in Hyperthermic, exercising individuals: a meta-analysis. J Athl Train. 2012;47(3):329–38.PubMedPubMedCentralCrossRef
38.
Zurück zum Zitat Casa DJ, Becker SM, Ganio MS, et al. Validity of devices that assess body temperature during outdoor exercise in the heat. J Athl Train. 2007;42(3):333–42.PubMedPubMedCentral Casa DJ, Becker SM, Ganio MS, et al. Validity of devices that assess body temperature during outdoor exercise in the heat. J Athl Train. 2007;42(3):333–42.PubMedPubMedCentral
39.
Zurück zum Zitat Purssell E, While A, Coomber B. Tympanic thermometry – normal temperature and reliability. Paediatr Nurs. 2009;21:40–3.PubMedCrossRef Purssell E, While A, Coomber B. Tympanic thermometry – normal temperature and reliability. Paediatr Nurs. 2009;21:40–3.PubMedCrossRef
Metadaten
Titel
Diagnostic test accuracy of new generation tympanic thermometry in children under different cutoffs: a systematic review and meta-analysis
verfasst von
Dan Shi
Li-Yuan Zhang
Hai-Xia Li
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Pediatrics / Ausgabe 1/2020
Elektronische ISSN: 1471-2431
DOI
https://doi.org/10.1186/s12887-020-02097-7

Weitere Artikel der Ausgabe 1/2020

BMC Pediatrics 1/2020 Zur Ausgabe

Mit dem Seitenschneider gegen das Reißverschluss-Malheur

03.06.2024 Urologische Notfallmedizin Nachrichten

Wer ihn je erlebt hat, wird ihn nicht vergessen: den Schmerz, den die beim Öffnen oder Schließen des Reißverschlusses am Hosenschlitz eingeklemmte Haut am Penis oder Skrotum verursacht. Eine neue Methode für rasche Abhilfe hat ein US-Team getestet.

Reanimation bei Kindern – besser vor Ort oder während Transport?

29.05.2024 Reanimation im Kindesalter Nachrichten

Zwar scheint es laut einer Studie aus den USA und Kanada bei der Reanimation von Kindern außerhalb einer Klinik keinen Unterschied für das Überleben zu machen, ob die Wiederbelebungsmaßnahmen während des Transports in die Klinik stattfinden oder vor Ort ausgeführt werden. Jedoch gibt es dabei einige Einschränkungen und eine wichtige Ausnahme.

Alter der Mutter beeinflusst Risiko für kongenitale Anomalie

28.05.2024 Kinder- und Jugendgynäkologie Nachrichten

Welchen Einfluss das Alter ihrer Mutter auf das Risiko hat, dass Kinder mit nicht chromosomal bedingter Malformation zur Welt kommen, hat eine ungarische Studie untersucht. Sie zeigt: Nicht nur fortgeschrittenes Alter ist riskant.

Begünstigt Bettruhe der Mutter doch das fetale Wachstum?

Ob ungeborene Kinder, die kleiner als die meisten Gleichaltrigen sind, schneller wachsen, wenn die Mutter sich mehr ausruht, wird diskutiert. Die Ergebnisse einer US-Studie sprechen dafür.

Update Pädiatrie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.