Anatomy of the coronary sinus with regard to cardiac resynchronization therapy implantation

verfasst von: Rasmus Borgquist, MD, PhD, Lingwei Wang, MD, PhD

Erschienen in: Herzschrittmachertherapie + Elektrophysiologie | Ausgabe 2/2022

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Knowledge of the coronary sinus (CS) anatomy is crucial for implantation of cardiac resynchronization therapy (CRT). Obstacles to CS entry, such as the Eustachian ridge and Thebesian valve, as well as within the CS, such as Vieussen’s valve and the vein of Marshall, are important to understand and differentiate during implantation or to identify earlier by imaging. Anatomic knowledge is mandatory to select the most suitable side branch for lead implantation. Modern tools and techniques almost always enable other anatomic problems, such as tortuous, small, short, or overly straight side branches, to also be overcome.
Hinweise
Scan QR code & read article online
Cardiac resynchronization therapy (CRT) is an established treatment for heart failure in selected patients and has formed part of the European Guidelines on cardiac pacing in heart failure for almost 20 years [1, 2]. The first case reports of biventricular pacing appeared in the late 1990s and showed promising results [35]. However, early attempts were hampered by the lack of specific devices tailored for biventricular pacing and lack of suitable tools for cannulation of the coronary sinus (CS), which allowed for both easy cannulation and the introduction of a left ventricular (LV) lead. Nevertheless, technical development was fast, and dedicated CRT pacemakers (CRT-P) and defibrillators (CRT-D) were introduced and approved by regulatory authorities. This set the stage for landmark prospective randomized trials, such as the MUSTIC, Care-HF, and COMPANION trials, which were later followed by the MADIT-CRT and RAFT trials [1, 69].
There are several technical challenges to overcome in order to achieve reliable LV pacing via the epicardial CS branches. Success rates have improved over time, as experience and technical tools have developed. In the early Care-HF study, the implant success rate at first attempt was 89%, and successful implants included 12% of the leads placed in the great cardiac vein (i.e., on the epicardial surface of the interventricular septum, thereby not directly stimulating the LV free wall) [10]. The reason for failed implant was primarily failure to cannulate the CS or failure to achieve a stable lead position, but also included perforation or dissection of the CS.
The introduction of modern bipolar or quadripolar leads, including active fixation, has made it possible to reach a stable position even in CS branches that are relatively short and therefore do not accommodate an LV lead with traditional curved passive fixation. Recently published experience from a large volume tertiary care center with experienced implanters reported an initial success rate of 99%, and active fixation leads had a lower dislodgement rate at 6 months (0.74% vs. 4.69%, p = 0.036) [11].
Successful CRT implants depend on a number of factors, related to either the implanting physician (such as implanter experience and skill), the implant tools, or patient characteristics (anatomic variants). This review focuses on the anatomical aspects of the CS, as seen from a CRT implanter perspective.

Normal anatomy

A thorough anatomical knowledge is paramount in order to achieve a high success rate for CS implants. The CS ostium is normally located in the lower portion of the posteroseptal aspect of the right atrium (Fig. 1).
The ostium lies between the septal aspect of the tricuspid valve, the fossa ovalis, and the inferior vena cava opening. On its way, it receives blood from several vein branches originating over the LV free wall, which are typically named after their location (i.e., posterolateral, lateral, anterolateral). The first branch (which can have a separate ostium) is usually the middle cardiac vein (MCV), which runs inferiorly in the interventricular groove to the apex of the heart (Fig. 2). The MCV may give off side branches that in theory can be used for lead placement, but usually these branches are very apical and thus not optimal for lead placement.
The main CS then runs posteriorly in the atrioventricular (AV) groove, and later becomes the great cardiac vein (GCV) at the level of Vieussen’s valve. It follows the AV groove to the anterior side of the heart and then runs on the superior aspect of the interventricular groove towards the apex, in parallel with the left anterior descending coronary artery (Fig. 3; [12]). The number of anatomically suitable vein branches traversing the LV between the MCV and the GCV varies, but on average there are between two and three potential vein branches that can be used [13].
The nomenclature of the vein branches can be somewhat confusing, especially since cardiologists have traditionally used the names posterior, posterolateral, lateral, anterolateral, and anterior for branches in the respective territories [14], whereas in echocardiography, cardiac computed tomography (CT) and cardiac magnetic resonance imaging (MRI), the traditional 17-segment model is used [15]. It is important to recognize that in order to determine the location of a pacing electrode tip in a particular cardiac segment, it is necessary to view two perpendicular X‑ray or fluoroscopy angles—the left anterior oblique (short axis view) and the right anterior oblique (long axis view) (Fig. 4). In the anteroposterior view alone, it is impossible to determine the location of a specific cardiac segment.

Anatomical challenges for the implanter

The anatomical challenges can be divided into the following parts:
1. CS access
i.
Right atrial enlargement
 
ii.
Thebesian valve
 
iii.
Anatomical variants
a.
CS stenosis
 
b.
Persistent left superior vena cava with or without CS atresia and unroofed CS
 
 
2. Vein-branch access
i.
Vieussen’s valve and the vein of Marshall
 
ii.
Separate ostium of posterior or posterolateral branch
 
iii.
Tortuous and small caliber veins
 
iv.
Phrenic nerve stimulation
 
1. CS access
i.
Before entering the CS ostium, the catheter or sheath needs to be advanced through the right atrium. With a severely enlarged right atrium, there is less or no support for the catheter if resistance is encountered at the CS ostium (Fig. 5). Therefore, the cardiac dimensions need to be reviewed prior to implant, and special tools may be required in the case of right atrial enlargement. Congenital disease such as Ebstein’s anomaly may also alter the location of the CS ostium. In patients with congenital malformations that could involve the atrium or venous system, a pre-operative cardiac CT is strongly recommended. An example is shown in Fig. 5, where there is extreme enlargement of the right atrium, but CT images show that there is still a patent entrance to the coronary sinus system in the usual location (Fig. 5).
 
ii.
The CS ostium may be partly, sometimes completely, covered by the Thebesian valve, which is a thin fibrous structure that may be fenestrated in some cases. Together with the extension of the Eustachian valve, the extension of the Thebesian valve forms a small ridge (Eustachian ridge) running in the right atrium, posterior to and in parallel with the tricuspid annulus. Together with the septal leaflet of the tricuspid valve and the extension of the inferior vena cava valve (the tendon of Todaro), the Eustachian ridge delineates the triangle of Koch, where the AV node is typically located. The ridge is derived from the valve of the embryonic sinus venosus (which is the origin of the valves of the CS and inferior vena cava). The Eustachian ridge may effectively prevent soft cannulation catheters from entering the CS, by deflecting them in a more posterior direction when trying to enter from the atrial side. This problem is best avoided by first entering the right ventricle with the catheter, and then retracting it during counter-clockwise torque. By using this technique, the ridge can actually be of help by guiding the catheter towards the CS ostium once it flips out from the right ventricle through the tricuspid valve.
 
iii.
The Thebesian valve can vary in configuration and texture. While it usually does not hamper CS cannulation, it may in some cases be very prominent and impossible to pass using a soft sheath. A prominent Thebesian valve can be seen on cardiac CT or suspected during implant if no or very little contrast enters the CS (Fig. 6). At times, a guidewire can pass through a fenestration, making it easy to advance the wire but impossible to advance the sheath after the wire, since the sheath becomes entangled in the valve around the fenestration. This problem can be overcome by using a steerable solid electrophysiology catheter either on the side or inside the soft sheath. The EP catheter can be manipulated to deflect the valve to allow for the soft sheath to advance into the CS.
 
iv.
Anatomical variants may affect the success rate of CS cannulation and lead placement. Narrowing of the CS ostium is not an uncommon finding and, in some cases, makes it impossible to pass with a sheath. This often occurs secondary to previous lead placements or other device placement in the CS, but can also occur without prior interventions. The problem can be overcome by either progressive balloon dilatation with a percutaneous coronary intervention (PCI) technique, or by using a stable steerable sheath that is positioned immediately outside the ostium, providing enough support for a sub-select catheter to enter. One alternative is to use a large 9‑F inner-diameter sheath outside the ostium and place a superstiff 0.035 in wire deep into the CS to keep the sheath stabilized and in place. Assistance in the CS cannulation via simultaneous femoral access with a diagnostic EP catheter may also increase the success rate of CS cannulation and lead placement in these difficult cases (Fig. 7).
Using the steerable catheter with real-time electrogram guidance during CRT implantation makes location of and access to the CS more rapid and successful, thus improving the success rate of CRT implantation and saving implantation and fluoroscopy times [16].
A persistent left superior vena cava (PLSVC) is the most common thoracic venous anomaly and may also be a component of complex cardiac pathologies. The prevalence of PLSVC ranges from 0.2 to 0.5% in the general population, but in patients with congenital heart disease between 1.3 and 11%. It may be accompanied by CS ostial atresia, in which case the PLSVC becomes the major retrograde drainage pathway for the coronary veins and there is secondary enlargement of the CS. In some cases, there is a direct connection between the CS and the left atrium (defined as an unroofed CS). In such a case, it is important to appreciate that this represents a potential right-to-left shunt and therefore permanent anticoagulation is indicated if a CS lead is placed. It is technically challenging to place an LV lead via a PLSVC, since the angle of take-off is usually steep and the enlarged main CS does not provide sufficient support. It is therefore recommended to use the right vena cava access for CS cannulation, unless there is CS atresia.
 
2. CS branch access
i.
Difficult ostium entrance. A common ostium for the MCV and the main CS can confuse the implanter as to where the sheath should be advanced. While the lateral branches from the main CS are usually preferred for LV lead placement, a side branch to the MCV can certainly be used as a second option in some cases. It is not uncommon that there is a tortuous take-off of the MCV from the common ostium, and cannulation can be challenging (Fig. 8).
By using slight clockwise torque from the ostium, it is often possible to pass a soft subselector or guidewire into the branch. If the side branch is short or without angle, lead stability may be an issue, and an active fixation lead is preferred.
 
ii.
Vieussen’s valve. The valve is situated where the main CS transforms into the GCV, which then continues to the superior part of the interventricular septum. It is composed of between one and three leaflets and is present in 80–90% of people. Normally, the leaflets are soft, non-occlusive, and easy to pass, but the valve can in rare cases be a challenging obstacle to pass with the sheaths and the electrode [17]. A sign of a prominent Vieussen’s valve is if contrast does not pass deeper into to CS when injected from a catheter from within the main CS trunk (Fig. 9). Immediately before the valve, there may be a patent vein of Marshall (the remnant distal part of the fetal left anterior cardinal vein), taking off in a superior direction and draining the left atrium. Vieussen’s valve may be associated with a curve of the CS, rendering its passage more difficult. Usually, it does not present a challenge for the CRT implanter, but there is the risk that the vein of Marshall is mistaken for the main CS trunk, and a balloon is inflated for an occlusive venogram. The vein of Marshall will not accommodate an inflated balloon, will burst and a tamponade may result if balloon inflation is attempted.
Using 4‑F catheters with a soft angulated tip in combination with hydrophilic wires, it is usually possible to get past the valve and then advance a subselect sheath further into a lateral branch. Often it is not necessary to pass the valve if there is a suitable posterolateral branch with take-off proximal to the valve.
 
iii.
Tortuous and small caliber vein branches are a common challenge for the implanter, and an array of tools may be needed to overcome this problem. Sometimes, the take-off is at an atypical angle, making it hard to engage the vein at all with a wire. Sometimes, the take-off is easy, but there is a tortuous course of the vein, making it impossible to advance sheaths or an electrode over the wire. Step one in this situation should always be to perform a venogram and make sure that there are no better options available in other branches. The use of a subselector sheath facilitates engagement of acute-angled vein branches, but is not always helpful if the tortuosity is more distal. Working with larger diameter subselectors allows for multiple guidewires to stabilize the sheath before advancing it. Other techniques involve snaring the guidewire through a collateral in the main CS (Fig. 10) or using the balloon anchoring technique. If the vein branch is too short to allow for proper anchoring of a passive curved fixation LV lead, active lead fixation may be used in the main CS, just letting the distal part protrude into the short vein branch.
 
iv.
The left phrenic nerve runs adjacent to the pericardium, and the exact course is unpredictable. Phrenic stimulation can occur in any of the lateral segments, and sometimes prevents delivering of LV pacing altogether. However, with the introduction of quadripolar leads, the problem has diminished considerably. By trying different vectors of stimulation, it is usually possible to find a configuration with capture that does not give rise to phrenic stimulation. In addition, most of the patients with CRT are not pacemaker-dependent, and low margins of safety for the LV lead can therefore be accepted if the threshold is stable.
 

Practical conclusion

This review presents the normal anatomy of the CS and its tributary vein branches from an electrophysiologist’s perspective. The most common anatomic variants and technical pitfalls for the implanting physician have been highlighted and concise guidance presented along with tips and tricks for overcoming these challenges. EP is a rapidly evolving field, and as more tools and implanting techniques become available, it is essential for every implanting physician to stay updated with the latest developments. A fundamental understanding of the cardiac anatomy is a prerequisite for becoming a successful CRT implanter.

Acknowledgments

The authors would like to thank Professor Elisabeth Englund and the staff at the pathology department for invaluable help in preparing the anatomical photos of explanted hearts.

Declarations

Conflict of interest

R. Borgquist declares speaker’s honoraria from Medtronic, Biotronik and Abbott. L. Wang declares that he has no competing interests.
For this article no studies with human participants or animals were performed by any of the authors. All studies mentioned were in accordance with the ethical standards indicated in each case.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

Herzschrittmachertherapie + Elektrophysiologie

Print-Titel

  • Themenbezogene Hefte zu aktuellen Fragen der Elektrophysiologie sowie der Diagnostik und Therapie von Herzrhythmusstörungen

  • Organ der Deutschen Gesellschaft für Kardiologie - Herz- und Kreislaufforschung e.V.
  • Journal affiliated to the European Heart Rhythm Association 
  • Partnerjournal der AG Rhythmologie der ÖKG

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Jetzt bestellen und 100 € sparen!

Literatur
1.
Zurück zum Zitat Cleland JG, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L et al (2005) The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 352(15):1539–1549CrossRef Cleland JG, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L et al (2005) The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 352(15):1539–1549CrossRef
2.
Zurück zum Zitat Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM et al (2021) 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J 42(35):3427–3520CrossRef Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM et al (2021) 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J 42(35):3427–3520CrossRef
3.
Zurück zum Zitat Barold SS, Cazeau S (2000) The first reports of electrical multisite ventricular activation in humans. Pacing Clin Electrophysiol 23(12):2117–2119CrossRef Barold SS, Cazeau S (2000) The first reports of electrical multisite ventricular activation in humans. Pacing Clin Electrophysiol 23(12):2117–2119CrossRef
4.
Zurück zum Zitat Le Franc P, Klug D, Lacroix D, Jarwe M, Kouakam C, Kacet S (1998) Triple chamber pacemaker for end-stage heart failure in a patient with a previously implanted automatic defibrillator. Pacing Clin Electrophysiol 21(8):1672–1675CrossRef Le Franc P, Klug D, Lacroix D, Jarwe M, Kouakam C, Kacet S (1998) Triple chamber pacemaker for end-stage heart failure in a patient with a previously implanted automatic defibrillator. Pacing Clin Electrophysiol 21(8):1672–1675CrossRef
5.
Zurück zum Zitat Cazeau S, Ritter P, Lazarus A, Gras D, Backdach H, Mundler O et al (1996) Multisite pacing for end-stage heart failure: early experience. Pacing Clin Electrophysiol 19(11 Pt 2):1748–1757CrossRef Cazeau S, Ritter P, Lazarus A, Gras D, Backdach H, Mundler O et al (1996) Multisite pacing for end-stage heart failure: early experience. Pacing Clin Electrophysiol 19(11 Pt 2):1748–1757CrossRef
6.
Zurück zum Zitat Linde C, Braunschweig F, Gadler F, Bailleul C, Daubert JC (2003) Long-term improvements in quality of life by biventricular pacing in patients with chronic heart failure: results from the Multisite Stimulation in Cardiomyopathy study (MUSTIC). Am J Cardiol 91(9):1090–1095CrossRef Linde C, Braunschweig F, Gadler F, Bailleul C, Daubert JC (2003) Long-term improvements in quality of life by biventricular pacing in patients with chronic heart failure: results from the Multisite Stimulation in Cardiomyopathy study (MUSTIC). Am J Cardiol 91(9):1090–1095CrossRef
7.
Zurück zum Zitat Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De Marco T et al (2004) Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 350(21):2140–2150CrossRef Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De Marco T et al (2004) Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 350(21):2140–2150CrossRef
8.
Zurück zum Zitat Moss AJ, Hall WJ, Cannom DS, Klein H, Brown MW, Daubert JP et al (2009) Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med 361(14):1329–1338CrossRef Moss AJ, Hall WJ, Cannom DS, Klein H, Brown MW, Daubert JP et al (2009) Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med 361(14):1329–1338CrossRef
9.
Zurück zum Zitat Goldenberg I, Kutyifa V, Klein HU, Cannom DS, Brown MW, Dan A et al (2014) Survival with cardiac-resynchronization therapy in mild heart failure. N Engl J Med 370(18):1694–1701CrossRef Goldenberg I, Kutyifa V, Klein HU, Cannom DS, Brown MW, Dan A et al (2014) Survival with cardiac-resynchronization therapy in mild heart failure. N Engl J Med 370(18):1694–1701CrossRef
10.
Zurück zum Zitat Gras D, Bocker D, Lunati M, Wellens HJ, Calvert M, Freemantle N et al (2007) Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: procedural success rate and safety. Europace 9(7):516–522CrossRef Gras D, Bocker D, Lunati M, Wellens HJ, Calvert M, Freemantle N et al (2007) Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: procedural success rate and safety. Europace 9(7):516–522CrossRef
11.
Zurück zum Zitat Robertson C, Duffey O, Tang PT, Fairhurst N, Monteiro C, Green P et al (2021) An active fixation quadripolar left ventricular lead for cardiac resynchronization therapy with reduced postoperative complication rates. J Cardiovasc Electrophysiol 33(3):458–463CrossRef Robertson C, Duffey O, Tang PT, Fairhurst N, Monteiro C, Green P et al (2021) An active fixation quadripolar left ventricular lead for cardiac resynchronization therapy with reduced postoperative complication rates. J Cardiovasc Electrophysiol 33(3):458–463CrossRef
12.
Zurück zum Zitat Noheria A, DeSimone CV, Lachman N, Edwards WD, Gami AS, Maleszewski JJ et al (2013) Anatomy of the coronary sinus and epicardial coronary venous system in 620 hearts: an electrophysiology perspective. J Cardiovasc Electrophysiol 24(1):1–6CrossRef Noheria A, DeSimone CV, Lachman N, Edwards WD, Gami AS, Maleszewski JJ et al (2013) Anatomy of the coronary sinus and epicardial coronary venous system in 620 hearts: an electrophysiology perspective. J Cardiovasc Electrophysiol 24(1):1–6CrossRef
13.
Zurück zum Zitat Markstad H, Bakos Z, Ostenfeld E, Geijer M, Carlsson M, Borgquist R (2019) Preoperative CT of cardiac veins for planning left ventricular lead placement in cardiac resynchronization therapy. Acta Radiol 60(7):859–865CrossRef Markstad H, Bakos Z, Ostenfeld E, Geijer M, Carlsson M, Borgquist R (2019) Preoperative CT of cardiac veins for planning left ventricular lead placement in cardiac resynchronization therapy. Acta Radiol 60(7):859–865CrossRef
14.
Zurück zum Zitat Singh JP, Klein HU, Huang DT, Reek S, Kuniss M, Quesada A et al (2011) Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial-cardiac resynchronization therapy (MADIT-CRT) trial. Circulation 123(11):1159–1166CrossRef Singh JP, Klein HU, Huang DT, Reek S, Kuniss M, Quesada A et al (2011) Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial-cardiac resynchronization therapy (MADIT-CRT) trial. Circulation 123(11):1159–1166CrossRef
15.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 105(4):539–542CrossRef Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 105(4):539–542CrossRef
16.
Zurück zum Zitat Wang L, Yuan S, Borgquist R, Hoijer CJ, Brandt J (2014) Coronary sinus cannulation with a steerable catheter during biventricular device implantation. Scand Cardiovasc J 48(1):41–46CrossRef Wang L, Yuan S, Borgquist R, Hoijer CJ, Brandt J (2014) Coronary sinus cannulation with a steerable catheter during biventricular device implantation. Scand Cardiovasc J 48(1):41–46CrossRef
17.
Zurück zum Zitat Strohmer B (2008) Valve of Vieussens: an obstacle for left ventricular lead placement. Can J Cardiol 24(9):e63CrossRef Strohmer B (2008) Valve of Vieussens: an obstacle for left ventricular lead placement. Can J Cardiol 24(9):e63CrossRef
Metadaten
Titel
Anatomy of the coronary sinus with regard to cardiac resynchronization therapy implantation
verfasst von
Rasmus Borgquist, MD, PhD
Lingwei Wang, MD, PhD
Publikationsdatum
01.06.2022
Verlag
Springer Medizin
Erschienen in
Herzschrittmachertherapie + Elektrophysiologie / Ausgabe 2/2022
Print ISSN: 0938-7412
Elektronische ISSN: 1435-1544
DOI
https://doi.org/10.1007/s00399-022-00863-x

Weitere Artikel der Ausgabe 2/2022

Herzschrittmachertherapie + Elektrophysiologie 2/2022 Zur Ausgabe

Mitteilungen aus der Arbeitsgruppe Elektrophysiologie und Rhythmologie (AGEP)

Mitteilungen aus der Arbeitsgruppe Elektrophysiologie und Rhythmologie (AGEP)

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.